Table 3.

Clinical performance characteristics of LFA in serum to detect SARS-CoV-2 IgG antibodies

IgG MedTest EcoTest Camtech
value 95% CI value 95% CI value 95% CI
Group PCR+/PCR–
 TP 59 63 59
 Sensitivity, % 78.67 67.68; 87.29 84.00 73.72; 91.45 78.67 67.68; 87.29
 Specificity, % 100.00 90.51; 100.00 89.19 74.58; 96.97 97.30 85.84; 99.93
 LR+ 30.39 4.37; 211.13 7.77 3.06; 19.71 29.11 4.20; 201.92
 LR– 0.21 0.14; 0.33 0.18 0.11; 0.30 0.22 0.14; 0.34
 DOR 144 31.21; 639 43.17 27.82; 65.70 132.32 30; 594
 Youden's index 0.79 0.58; 0.87 0.73 0.48; 0.88 0.76 0.54; 0.87
 CUI+ 0.79 0.70; 0.88 0.79 0.70; 0.88 0.77 0.68; 0.87
 CUI– 0.70 0.61; 0.79 0.65 0.55; 0.76 0.67 0.58; 0.77
 PPV, % 100.00 94.03 86.13; 97.56 98.33 89.48; 99.76
 NPV, % 69.81 59.96; 78.12 73.33 61.80; 82.38 69.23 59.22; 77.71
 Accuracy, % 85.71 77.84; 91.61 85.71 77.84; 91.61 84.82 76.81; 90.90
 MCC 0.74 0.70 0.72
Group PCR+
 TP 58 62 57
 Sensitivity, % 87.88 77.51; 94.62 93.94 85.20; 98.32 86.36 75.69; 93.57
 Specificity, % 100.00 90.51; 100.00 89.19 74.58; 96.97 97.30 85.84; 99.93
 LR+ 33.84 4.88; 234.75 8.69 3.44; 21.97 31.95 4.61; 221.41
 LR– 0.12 0.06; 0.23 0.07 0.03; 0.18 0.14 0.08; 0.26
 DOR 282 81.33; 1,020.00 124 115; 122 228 7.62; 850
 Youden's index 0.88 0.68; 0.95 0.83 0.60; 0.95 0.84 0.62; 0.94
 CUI+ 0.88 0.81; 0.95 0.88 0.81; 0.95 0.85 0.77; 0.93
 CUI– 0.82 0.75; 0.90 0.80 0.71; 0.88 0.78 0.70; 0.86
 PPV, % 100.00 93.94 85.98; 97.51 98.28 89.16; 99.75
 NPV, % 82.22 70.72; 89.85 89.19 76.02; 95.55 80.00 68.50; 88.04
 Accuracy, % 92.23 85.27; 96.59 92.23 85.27; 96.59 90.29 82.87; 95.25
 MCC 0.85 0.83 0.81
Group PCR–
 TP 01 01 02
 Sensitivity, % 11.11 0.28; 48.25 11.11 0.28; 48.25 22.22 2.81; 60.01
 Specificity, % 100.00 90.51; 100.00 89.19 74.58; 96.97 97.30 85.84; 99.93
 LR+ 7.09 0.71; 71.11 1.03 0.13; 8.12 8.22 0.83; 80.97
 LR– 0.89 0.71; 1.12 1.00 0.77; 1.29 0.80 0.56; 1.14
 DOR 7.97 1.00; 63.49 1.03 0.17; 6.30 10.28 1.48; 71.03
 Youden's index 0.11 −0.09; 0.48 0.003 −0.25; 0.45 0.20 −0.11; 0.60
 CUI+ 0.11 0.00; 0.65 0.02 0.00; 0.35 0.15 0.00; 0.64
 CUI– 0.82 0.75; 0.90 0.72 0.62; 0.81 0.82 0.74; 0.89
 PPV, % 100.00 20.00 3.07; 66.38 66.67 16.88; 95.17
 NPV, % 82.22 78.59; 85.35 80.49 76.14; 84.21 83.72 78.32; 87.98
 Accuracy, % 82.61 68.58; 92.18 73.91 58.87; 85.73 82.61 68.58; 92.18
 MCC 0.30 0.003 0.31

LFA, lateral flow immunochromatographic assay; TP, true positive; PPV, positive predictive value; NPV, negative predictive value; LR, likelihood ratio; DOR, diagnostic odds ratio (the higher the DOR value, the better the test); CUI, clinical utility index; MCC, Matthews correlation coefficient.