Table 6.
Detection performance of N1 and N2 UltraPlex assay-probe sets on heat-treated swab samples, compared to Abbott assay
Probe |
Results (n) |
Accuracy (%) |
Avg CT shift±sem |
|||
---|---|---|---|---|---|---|
TP |
TN |
FP |
FN |
|||
N1 |
95 |
18 |
2 |
12 |
88.98 |
12.54±0.29 |
N2 |
98 |
20 |
0 |
9 |
92.91 |
13.52±0.18 |
TP, true positive; TN, true negative; FP, false positive; FN, false negative.