Table 1.

Intra-day and inter-day precision and accuracy of LC/MS/MS method for the determination of erinacine A.

Intra-Day Inter-Day
Theoretical Conc. Observed Conc. Precision Accuracy Observed Conc. Precision Accuracy
(ng/mL) (ng/mL) (% CV) (% bias) (ng/mL) (% CV) (% bias)
5 4.27 ± 0.37 8.79 −14.70 4.08 ± 0.52 9.34 −9.66
10 12.85 ± 0.68 8.23 14.00 11.40 ± 0.94 4.04 10.33
20 17.48 ± 0.61 3.50 −12.58 16.99 ± 0.98 5.76 −15.00
50 52.55 ± 1.04 1.98 5.10 49.69 ± 1.28 2.59 −0.62
100 104.50 ± 1.76 1.68 4.50 103.49 ± 12.03 11.63 3.49
200 207.83 ± 2.40 1.16 3.92 178.67 ± 9.19 5.14 −10.67
500 479.17 ± 8.84 1.85 −4.17 466.33 ± 31.11 6.67 −6.73

Data expressed as means ± SD (n = 6). CV (%) = (standard deviation/mean) × 100%. bias (%) = [(measured concentration − theoretical concentration)/theoretical concentration] × 100%.