. 2021 Jul 24;10(15):3274. doi: 10.3390/jcm10153274

Table 1.

Imaging parameters of the reference prospectively triggered steady-state free-precession cine imaging and real-time compressed sensing cine imaging.

Parameters Cineref CSrt
Repetition time—ms 3.16 2.70
Echo time—ms 1.23 1.14
Flip angle—degrees 57 60
Field of view—mm2 375 × 280 360 × 270
Matrix—pixels2 288 × 216 224 × 168
Spatial resolution—mm2 1.3 × 1.3 1.6 × 1.6
Temporal resolution—ms 41.2 49
Slice thickness/gap—mm 8/2 8/2
Bandwidth—Hz/pixel 915 900
ECG mode Prospective triggering Adaptative triggering
Number of measured cardiac phases per cycle 20 a 17.0 ± 3.2
Number of reconstructed cardiac frames per cycle—n 20 a 20 b
Number of views per frame—n 13.0 ± 4.8 c 18 a
Cycles of iterative reconstruction—n NA 40
Acceleration factor 2 11

Data are expressed as mean ± standard deviation in the absence of any indication. a Constant value. b Interpolation was performed to provide a constant frame rate of 20 cardiac phases per cycle for post-processing. c The number of views per frame was set according to the shorter RR interval in order to acquire 20 cardiac phases. Prospective triggering allows data sampling during a fixed acquisition window after each R peak while adaptative triggering allow data sampling until the next R peak occurs. Abbreviations: Cineref, reference segmented cine; CSrt, real-time compressed sensing cine; ECG, electrocardiogram; n, data represented as numbers; NA, not applicable.