Abstract
Purpose
High resolution flat-panel computed tomography arthrography (FPCT-A) and magnetic resonance arthrography (MR-A) are well suited to evaluate osteochondral lesions. The current study compares the performance of FPCT-A versus MR-A in an experimental setting.
Methods
Fourteen cadaveric ankles were prepared with artificial osteochondral defects of various sizes in four separate talar locations. After intra-articular contrast injection, FPCT-A and 3-T MR-A were acquired. Each defect was then filled with synthetic pallets. The resulting cast was used as reference. Two independent radiologists measured the dimensions of all defects with FPCT-A and MR-A. Intra-class correlation coefficients (ICC) were calculated. Data were compared using t-tests and Bland-Altman plots.
Results
The correlation for FPCT-A and cast was higher compared to MR-A and cast (ICC 0.876 vs. 0.799 for surface [length x width]; ICC 0.887 vs. 0.866 for depth, p<0.001). Mean differences between FPCT-A and cast measurements were -1.1 mm for length (p<0.001), -0.7 mm for width (p<0.001) and -0.4 mm for depth (p = 0.023). By MR-A, there were no significant differences for length and width compared to cast (p>0.05). Depth measurements were significantly smaller by MR-A (mean difference -1.1 mm, p<0.001). There was no bias between the different modalities.
Conclusions
Ex vivo FPCT-A and MR-A both deliver high diagnostic accuracy for the evaluation of osteochondral defects. FPCT-A was slightly more accurate than MR-A, which was most significant when measuring lesion depth.
Introduction
At the ankle, osteochondral lesions are often traumatic and painful [1, 2]. The lesions typically involve the articular cartilage and subchondral bone, mostly the talus. Surgery is required in many cases. The optimal form of surgical treatment depends on defect characteristics, such as size and location [3–7]. Further, defect size correlates with patient outcome [8]. Therefore, it is imperative to be able to precisely determine the extent of any osteochondral defects at the ankle [9].
Routine magnetic resonance imaging (MRI) without intravenous or intra-articular contrast agent is an option to evaluate the cartilage at the ankle [10–12]. Standard MRI, however, may be limited and overestimate the size of a lesion due to bone edema [13]. Direct magnetic resonance arthrography (MR-A) could be considered for further work up. MR-A better detects chondral defects, especially in patients with lesions of the upper ankle cartilage [14].
Improved dose reduction management and technology in multidetector computed tomography (MDCT) have enabled MDCT to also detect thin cartilaginous lesions at the ankle [15]. Recently, flat panel computed tomography (FPCT) was established as an imaging modality with superior spatial resolution including submillimeter isotropic voxel sizes compared to MDCT [16, 17]. The radiation dose is similar for FPCT and MDCT [17]. FPCT further presents reduced metal and beam-hardening artifacts [18]. Finally, FPCT is widely available as a part of most angiography suites. There, fluoroscopic guided contrast injection and cross-sectional arthrography can be performed in a single session without need for patient transfer [19, 20].
Despite the plethora of various imaging techniques, a definitive diagnosis of an osteochondral defect can be challenging [18, 21]. Controversy exists regarding the imaging modality of choice [2, 22–24]. Thus, the current study investigates and compares the diagnostic performance of high resolution FPCT-A versus MR-A in an experimental setup.
Materials and methods
Cadaver preparation
The study was waived by the local Institutional Review Board (Medical School Hannover). Fourteen fresh-frozen cadaveric feet with lower legs were recruited for the study (Science Care). Any prior surgical treatment resulted in exclusion. The upper ankle joint was opened by a mini-arthrotomy and inspected for already existing osteochondral lesions. Only cadavers without any defects were used. Artificial defects of different sizes and depths were then created with different surgical curettes. The lesions were established in four predefined locations at the talus: anteromedial, anterolateral, posteromedial and posterolateral. Collateral soft tissue damage was avoided during the procedure. Then, the joint capsule and skin were closed with surgical sutures.
Contrast application and image acquisition
The cadavers were positioned with the lateral malleolus facing upwards. The injection was performed under fluoroscopic guidance using a 20G needle. The contrast consisted of a 1:1 mixture of an iodinated agent (Iomeprol 300 mg/ml, Imeron 300, Bracco Imaging) and a pre-diluted gadolinium-based solution (Gadolinium-DTPA 2 mmol/l, Dotarem 0,5 mmol/ml, Guerbet). 5–10 ml contrast mixture was injected into each joint. Then, the ankle was passively moved to ensure homogeneous distribution of the contrast agent within the joint.
Following the injection, FPCT-A was acquired with the same angiographic system (ARTIS pheno AXH 1964, Siemens Healthineers). Tube voltage was 70 kV, tube current was 400–410 mA. The average dose-area product was 685 μGym2. Images were reconstructed using the smallest possible volume-of-interest per ankle and the “normal” bone reconstruction kernel. The resulting voxel size was in the range of 0.11–0.12 mm.
After transfer to the MRI suite, MR images were acquired using a 3T MR scanner (MAGNETOM Skyra, Siemens Healthineers) and a dedicated 20-channel phase array head coil. As imaging protocol, the protocol already established for clinical application was used comprised of the following: T1-weighted (w) Turbo Spin Echo (TSE) sequences with spectral fat suppression (FS) in three planes; sagittal T1w TSE sequence without FS; a proton density weighted (PDw) coronal sequence; an isotropic 3D PD space sequence. The detailed scan parameters are summarized in Table 1.
Table 1. MR scan parameters.
| MR-A | T1w TSE | T1w TSE | T1w TSE | PDw TSE | T1w TSE | 3DTSE SPACE |
|---|---|---|---|---|---|---|
| Parameters | FS | FS | FS | FS | FS | |
| Orientation | transverse | sagittal | sagittal | coronal | coronal | sagittal |
| 3D block | ||||||
| Repetition time (ms) | 601 | 956 | 1010 | 2660 | 721 | 1000 |
| Echo time (ms) | 13 | 13 | 13 | 33 | 13 | 43 |
| Receiver bandwidth | 257 | 257 | 257 | 200 | 257 | 460 |
| (Hertz/pixel) | ||||||
| Flip angle (degree) | 150 | 150 | 150 | 150 | 150 | variable |
| Field-of-view (mm2) | 130 x 110 | 130 x 130 | 130 x 130 | 130 x 102 | 130 x 130 | 180 x 170 |
| Matrix | 512 x 367 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 320 x 285 |
| Slice thickness / gap (mm) | 2.0/0 | 2.0/0 | 2.0/0 | 2.0/0 | 2.0/0 | 0.6/0 |
| Pixel dimensions (mm2) | 0.3 x 0.3 | 0.3 x 0.3 | 0.3 x 0.3 | 0.3 x 0.3 | 0.3 x 0.3 | 0.6 x 0.6 |
| Number of Slices | 30 | 32 | 30 | 40 | 33 | 112 |
| Phase encoding direction | anterior to posterior | head to foot | head to foot | right to left | right to left | head to foot |
| Acquisition time (min:sec) | 4:02 | 4:22 | 4:40 | 3:08 | 3:45 | 7:26 |
Magnetic Resonance Arthrography (MR-A) scanning parameters
TSE = turbo spin echo; FS = fat saturated; PDw = proton density weighted; space = sampling perfection with application optimized contrasts using different flip angle evolution
Preparation of casting specimens
After arthrography, the cadaveric ankles were disarticulated and the talar surface was carefully prepared. In order to obtain a reference size for each lesion, the defects were filled with synthetic pallets (Friendly plastic, AMACO). This created a reference cast for each defect. The casts were labeled according to the defect location and a high-resolution FPCT-scan was acquired for measurements of the casts.
Image assessment and measurements
Two independent radiologists with more than 5 years in practice evaluated all FPCT-A and MR-A images. The radiologists were blinded to the identifying information of the images. Images were evaluated on a client-server based picture archiving and communication system (PACS) workstation (Visage 7.1, Visage Imaging). The raters used the 3D functionality of the software including multiplanar reformations, thin slab averaging and fusion/overlay mode. Regarding the MR-images, the radiologists evaluated defect configuration on all MR sequences and used the 3D SPACE sequence in the multiplanar mode to measure the dimensions of the defects. Defect depth was measured as illustrated in Fig 1A. Defect length was defined as the largest extent at the cartilage level, defect width was measured at the largest extent of a perpendicular line to the defect’s length (Fig 1A). Defect length and width were multiplied and defined as the defect’s surface area (mm2) and used for calculation of correlation coefficients and Bland-Altman plots. Measurement of the casting specimens included visualization in multiplane reconstruction and 3D mode with a clear depiction of defect shape. Then, the cast’s length, width and depth were identified and measured in three planes. The results were calculated similarly to the osteochondral defect measurements (Fig 1B).
Fig 1. Osteochondral defect measurement.
Example of measurements in flat panel computed tomography arthrography (FPCT-A) with a multiplane reconstruction of the same defect in three planes (A) and the corresponding cast specimen (B) in two planes and a 3D rendering mode.
Statistical analysis
SPSS Statistics 27.0.0.0 (IBM Corp) was used for the statistical analysis. Tests on normal distribution (Kolmogorov-Smirnow) for both readers and all FPCT-, MR-A and cast specimen measurements were performed. The defect measurements of reader 1 were compared to reader 2 using a t-test. The inter-reader correlation was calculated using intra-class correlation coefficient (ICC) with associated p value. The means of corresponding measurements of the two radiologists were calculated and used for comparison/correlation of FPCT-A, MR-A and cast.
Reliability between both imaging modalities and casting specimens were calculated using ICCs with a 95% confidence interval in a two-way, mixed, consistency model [25]. The ICC values were interpreted as follows: values < 0.4 reflected poor correlation, 0.4–0.59 intermediate, 0.60–0.74 good, and 0.75–1.00 very good correlation [26]. Test on normal distribution (Kolmogorov-Smirnow) showed normal distribution for all FPCT-, MR-A and cast specimen measurements. A t-test for paired samples was used for comparison of the measurements of FPCT-A vs. cast, MR-A vs. cast and FPCT-A vs. MR-A. A p-value < 0.05 was considered significant. Bland-Altman plots were calculated. The upper and lower limit of agreement range was calculated according to: mean difference +/- standard deviation x 1.96.
Results
All four artificially created defects in each ankle joint were detected by both the FPCT- and MR-images. The defect measurements of reader 1 compared to reader 2 showed no significant difference (p>0.05). Indeed, the inter-reader correlation between reader 1 and 2 showed “very good” agreement (ICC >0.9, 95% CI 0.984–0.989, p<0.001). The mean values of corresponding measurements of the two radiologists are presented in Tables 2 and 3.
Table 2. Defect measurements cadavers 1–7.
| Cadaver No. | Talar defect location | Surface in mm2 | Depth in mm | ||||
|---|---|---|---|---|---|---|---|
| mean of both readers (SD) | mean of both readers (SD) | ||||||
| FPCT-A | MR-A | Cast | FPCT-A | MR-A | Cast | ||
| 1 | AM | 28.8 (0.0) | 41.4 (7.8) | 32.9 (3.1) | 4.9 (0.1) | 4.8 (0.0) | 4.6 (0.1) |
| AL | 47.2 (1.4) | 162.5 (46.32) | 34.4 (11.9) | 3.6 (0.0) | 3.5 (0.0) | 2.1 (0.1) | |
| PM | 45.7 (1.5) | 50.7 (1.6) | 52.0 (7.9) | 14.3 (0.3) | 14.0 (0.0) | 9.6 (0.0) | |
| PL | 38.9 (0.0) | 41.0 (2.4) | 55.4 (3.3) | 5.3 (0.0) | 4.0 (0.0) | 3.9 (0.1) | |
| 2 | AM | 76.8 (0.0) | 93.3 (0.4) | 75.8 (21.4) | 6.4 (0.0) | 5.4 (0.1) | 8.2 (0.1) |
| AL | 37.5 (0.0) | 43.4 (0.9) | 36.6 (9.5) | 5.7 (0.0) | 4.7 (0.1) | 5.8 (0.1) | |
| PM | 42.4 (0.0) | 44.8 (0.4) | 42.3 (6.4) | 6.8 (0.0) | 4.9 (0.0) | 5.0 (1.8) | |
| PL | 65.1 (0.0) | 68.9 (2.3) | 65.98(7.67) | 6.6 (0.0) | 4.8 (0.1) | 6.9 (2.7) | |
| 3 | AM | 73.9 (0.0) | 92.4 (0.9) | 70.7 (15.8) | 4.0 (0.0) | 4.0 (0.1) | 4.9 (1.1) |
| AL | 42.6 (0.0) | 51.2 (4.0) | 77.04 (0.76) | 1.6 (0.0) | 1.5 (0.0) | 3.2 (0.8) | |
| PM | 65.5 (0.0) | 69.7 (1.2) | 73.6 (11.2) | 5.5 (0.0) | 4.0 (0.1) | 5.3 (0.8) | |
| PL | 41.6 (0.0) | 41.1 (3.9) | 48.8 (13.7) | 2.7 (0.0) | 2.5 (0.0) | 2.7 (0.4) | |
| 4 | AM | 34.8 (0.0) | 48.6 (0.5) | 92.3 (44.7) | 1.4 (0.1) | 1.3 (0.1) | 2.0 (1.1) |
| AL | 52.9 (0.5) | 66.0 (1.7) | 76.0 (6.0) | 4.2 (0.1) | 2.6 (0.1) | 4.2 (0.6) | |
| PM | 72.7 (1.4) | 78.2 (1.3) | 90.0 (7.0) | 2.4 (0.3) | 2.1 (0.1) | 6.4 (0.4) | |
| PL | 36.3 (11.0) | 51.6 (0.0) | 55.8 (9.0) | 4.0 (0.9) | 2.2 (0.0) | 4.5 (0.5) | |
| 5 | AM | 97.1 (0.5) | 106.5 (2.9) | 138.8 (9.4) | 6.1 (0.0) | 3.2 (0.1) | 4.7 (0.4) |
| AL | 38.7 (3.8) | 48.2 (0.4) | 75.5 (9.2) | 4.5 (0.1) | 3.8 (0.3) | 4.2 (0.6) | |
| PM | 93.5 (0.7) | 98.2 (0.7) | 89.3 (6.3) | 5.4 (0.1) | 5.4 (0.1) | 6.8 (0.8) | |
| PL | 36.0 (0.0) | 64.3 (4.7) | 63.8 (2.0) | 5.1 (0.1) | 3.3 (0.1) | 4.2 (0.6) | |
| 6 | AM | 92.9 (8.5) | 122.3 (0.7) | 114.2 (24.2) | 5.9 (0.2) | 5.4 (0.1) | 5.6 (0.1) |
| AL | 61.8 (0.5) | 87.7 (0.7) | 94.9 (13.7) | 1.6 (0.0) | 1.6 (0.1) | 2.9 (0.4) | |
| PM | 79.6 (1.97) | 139.1 (14.0) | 92.4 (26.5) | 5.0 (0.1) | 4.8 (0.0) | 6.2 (0.1) | |
| PL | 42.6 (2.3) | 57.5 (0.1) | 62.8 (2.6) | 4.9 (0.1) | 3.8 (0.1) | 7.2 (0.1) | |
| 7 | AM | 112.7 (2.4) | 119.5 (0.0) | 89.6 (2.2) | 5.7 (0.1) | 4.7 (0.1) | 6.6 (0.0) |
| AL | 35.6 (0.1) | 66.8 (0.6) | 74.6 (0.0) | 1.2 (0.1) | 1.1 (0.1) | 2.0 (0.5) | |
| PM | 34.8 (1.7) | 66.0 (0.7) | 44.7 (4.7) | 1.3 (0.1) | 2.4 (0.1) | 2.6 (0.0) | |
| PL | 74.1 (0.8) | 119.9 (1.5) | 71.9 (10.9) | 3.2 (0.1) | 4.4 (0.2) | 4.1 (0.2) | |
Mean values of both readers for each localization in cadavers1-7 for surface and depth measurements in FPCT-A, MR-A and cast respectively. Standard deviation (SD) in parantheses. AM = anteromedial, AL = anterolateral, PM = posteromedial and PL = posterolateral.
Table 3. Defect measurements cadavers 8–14.
| Cadaver No. | Talar defect location | Surface in mm2 | Depth in mm | ||||
|---|---|---|---|---|---|---|---|
| mean of both readers (SD) | mean of both readers (SD) | ||||||
| FPCT-A | MR-A | Cast | FPCT-A | MR-A | Cast | ||
| 8 | AM | 83.0 (0.0) | 97.1 (2.7) | 84.2 (17.5) | 1.5 (0.0) | 1.6 (0.1) | 2.9 (0.3) |
| AL | 81.8 (1.1) | 119.8 (3.1) | 104.4 (6.6) | 3.4 (0.1) | 3.1 (0.1) | 4.6 (0.4) | |
| PM | 77.9 (0.2) | 93.2 (8.8) | 100.5 (0.7) | 4.5 (0.0) | 3.9 (0.2) | 5.4 (0.3) | |
| PL | 36.5 (0.0) | 36.3 (0.4) | 61.0 (2.7) | 1.5 (0.1) | 1.4 (0.0) | 2.2 (0.2) | |
| 9 | AM | 89.9 (0.0) | 76.3 (3.6) | 87.7 (12.6) | 2.9 (0.0) | 1.8 (0.0) | 3.8 (0.3) |
| AL | 83.7 (0.0) | 109.7 (34.6) | 97.6 (8.8) | 3.4 (0.0) | 2.3 (0.5) | 3.3 (0.0) | |
| PM | 97.5 (0.0) | 56.0 (3.5) | 94.4 (24.0) | 5.1 (0.0) | 2.2 (0.3) | 5.7 (0.8) | |
| PL | 75.6 (0.0) | 85.5 (3.5) | 108.5 (0.2) | 4.2 (0.0) | 2.5 (0.0) | 5.0 (0.7) | |
| 10 | AM | 77.8 (2.5) | 120.1 (20.3) | 105.4 (14.7) | 4.5 (0.1) | 3.3 (0.1) | 4.3 (0.4) |
| AL | 115.5 (6.2) | 100.4 (4.94) | 162.7 (3.8) | 5.4 (0.1) | 4.9 (0.0) | 6.3 (0.5) | |
| PM | 84.2 (0.7) | 93.1 (1.4) | 82.3 (6.6) | 7.8 (0.2) | 6.7 (0.4) | 7.4 (0.5) | |
| PL | 82.5 (9.2) | 81.4 (5.9) | 95.9 (16.9) | 2.0 (0.0) | 1.8 (0.1) | 2.6 (0.1) | |
| 11 | AM | 163.1 (17.5) | 275.5 (0.0) | 200.2 (24.5) | 6.3 (0.1) | 5.4 (0.0) | 6.2 (0.0) |
| AL | 120.2 (2.6) | 104.2 (36.4) | 94.9 (12.5) | 1.1 (0.1) | 1.4 (0.2) | 2.5 (0.4) | |
| PM | 197.6 (3.7) | 96.0 (8.5) | 127.3 (13.6) | 6.4 (0.1) | 6.3 (0.1) | 7.5 (0.1) | |
| PL | 83.2 (0.1) | 93.5 (2.0) | 107.3 (5.6) | 2.7 (0.2) | 2.3 (0.3) | 4.5 (0.5) | |
| 12 | AM | 110.3 (1.8) | 175.2 (2.7) | 156.3 (54.2) | 4.8 (0.1) | 3.3 (0.1) | 4.0 (0.7) |
| AL | 85.64 (1.3) | 86.5 (4.4) | 105.6 (16.2) | 4.1 (0.1) | 4.0 (0.1) | 5.5 (0.0) | |
| PM | 77.0 (1.4) | 71.8 (2.7) | 103.9 (2.8) | 1.6 (0.0) | 1.7 (0.1) | 2.7 (0.1) | |
| PL | 43.5 (0.0) | 67.6 (0.7) | 74.6 (10.3) | 4.7 (0.1) | 4.8 (0.1) | 4.4 (0.7) | |
| 13 | AM | 99.4 (0.0) | 129.5 (2.0) | 115.0 (33.6) | 3.2 (0.2) | 3.4 (0.1) | 3.3 (0.1) |
| AL | 50.1 (1.6) | 42.0 (12.8) | 109.7 (12.4) | 3.7 (0.1) | 2.7 (0.1) | 3.7 (0.1) | |
| PM | 58.0 (0.5) | 71.8 (1.8) | 60.9 (5.7) | 4.3 (0.4) | 3.9 (0.1) | 5.5 (0.6) | |
| PL | 52.3 (4.6) | 43.4 (0.1) | 65.1 (4.5) | 4.4 (0.1) | 4.3 (0.1) | 4.0 (1.5) | |
| 14 | AM | 82.7 (0.0) | 88.7 (0.2) | 83.5 (12.9) | 4.3 (0.0) | 2.6 (0.1) | 4.1 (0.8) |
| AL | 111.0 (0.0) | 131.8 (3.2) | 152.2 (14.6) | 5.6 (0.0) | 3.7 (0.3) | 6.1 (1.6) | |
| PM | 91.0 (0.0) | 103.3 (5.0) | 103.0 (5.1) | 3.3 (0.0) | 2.1 (0.1) | 3.7 (0.6) | |
| PL | 65.3 (0.0) | 71.1 (11.2) | 70.5 (0.4) | 5.3 (0.0) | 4.7 (0.2) | 5.3 (0.4) | |
Mean values of both readers for each localization in cadavers 8–14 for surface and depth measurements in FPCT-a, MR-A and cast respectively. Standard deviation (SD) in parantheses. AM = anteromedial, AL = anterolateral, PM = posteromedial and PL = posterolateral.
T-test results
The defects were measured significantly smaller compared to the cast using FPCT-A: mean differences were -1.1 mm for length (p<0.001), -0.7 mm for width (p<0.001) and -0.4 mm for depth (p = 0.023). By MR-A, there was no significant difference for length and width compared to the cast, whereas depth was measured significantly smaller in MR-A (mean difference -1.1 mm, p<0.001). Overall, defect width and length were measured significantly smaller by FPCT-A compared to MR-A (p<0.001), whereas depth was measured +0.7 mm larger in FPCT-A (p<0.001). The detailed results are given in Table 4.
Table 4. T-test results.
| Defect size parameter | Modalities | Mean difference | [95% CI] | t-value | p-value |
|---|---|---|---|---|---|
| Length | FPCT-A–cast | -1.1 | [-1.7; -0.5] | -3.8 | <0.001 |
| (mm) | |||||
| MR-A–cast | -0.6 | [-1.2; 0.0] | -1.9 | 0.062 | |
| FPCT-A–MR-A | -0.5 | [-1.1; 0.0] | -2.0 | 0.056 | |
| Width | FPCT-A–cast | -0.7 | [-1.0; -0.4] | -4.3 | <0.001 |
| (mm) | |||||
| MR-A–cast | 0.2 | [0.6; 1.1] | 1.1 | 0.289 | |
| FPCT-A–MR-A | -0.9 | [-1.2; -0.6] | -5.7 | <0.001 | |
| Depth | FPCT-A–cast | -0.4 | [-0.7; -0.1] | -2.3 | 0.023 |
| (mm) | |||||
| MR-A–cast | -1.1 | [-1.4; -0.7] | -6.4 | <0.001 | |
| FPCT-A–MR-A | 0.7 | [0.5; 0.9] | 6.2 | <0.001 |
Results of the paired t-test with mean difference of defect length, width and depth (in mm) between the flat panel computed tomography arthrography (FPCT-A), magnetic resonance arthrography (MR-A) and cast specimens. Data are presented with 95% confidence intervals (CI), t-values and p-values.
Correlation results
All calculated ICCs between FPCT-A, MR-A and cast showed very good correlation (ICC >0.75, p<0.001). The measurements of the defect surface (length x width) showed a higher correlation for FPCT-A and cast (ICC 0.876, p<0.001) compared to MR-A and cast (ICC 0.799, p<0.001). Similarly, depth measurements correlated slightly better between FPCT-A and cast (0.887, p<0.001) compared to MR-A and cast (0.866, p<0.001). We observed a high correlation of the depth measurements between FPCT-A and MR-A (0.957, p<0.001). The detailed results are given in Table 5.
Table 5. Correlation results.
| Defect size parameter | Modalities | ICC | 95% CI | p-value |
|---|---|---|---|---|
| Surface (mm 2 ) | FPCT-A vs. cast | 0.876 | [0.788; 0.927] | <0.001 |
| MR-A vs. cast | 0.799 | [0.656; 0.882] | <0.001 | |
| FPCT-A vs. MR-A | 0.796 | [0.652; 0.880] | <0.001 | |
| Depth | FPCT-A vs. cast | 0.887 | [0.808; 0.934] | <0.001 |
| (mm) | MR-A vs. cast | 0.866 | [0.771; 0.921] | <0.001 |
| FPCT-A vs. MR-A | 0.957 | [0.927; 0.975] | <0.001 |
Intra-class correlation coefficients (ICC) for surface (in mm2) and depth (in mm) between flat panel computed tomography arthrography (FPCT-A), magnetic resonance arthrography (MR-A) and cast specimens. Data are presented with 95% confidence intervals (CI) and p-values
Bland-Altman analysis
The Bland-Altman plots (Fig 2) demonstrate the mean differences of the compared imaging modalities and cast with almost all values within the limit of agreement range. There was no bias between FPCT-A, MR-A and cast.
Fig 2. Bland-Altman plots.

Bland-Altman plots for flat panel computed tomography arthrography (FPCT-A), magnetic resonance arthrography (MR-A) and cast specimens with mean differences of surface (A) and depth (B) measurements (red line) and ± 1.96*standard deviation (green dotted lines). The x-axis describes the mean of the data and the y-axis shows the difference of the data.
Image differences of spatial resolution between FPCT-A and MR-A are demonstrated in Fig 3 with a single defect reconstructed in two corresponding planes in MR-A (A) and FPCT-A (B) respectively. Fig 4 shows another single defect in the PD TSE FS (A) and T1 TSE (C) MR-sequences with the corresponding FPCT-A planes (B and D). Fig 5 represents a fusion image of an osteochondral talar defect in FPCT-A with an overlay FPCT-image of the corresponding cast specimen.
Fig 3. 3D MR- and flat panel CT-arthrography measurements.
Multiplane reconstruction of one single defect in the isotropic 3D space MR-sequence (A) with the corresponding planes in flat panel computed tomography arthrography (B) with marked depth measurements. The margins of the artificially created defect are clearly defined against the intra-articular contrast agent and the defect dimensions can be measured.
Fig 4. Different MR sequences and flat panel CT-arthrography measurements.
Proton density weighted (PDw) coronal (A) and T1w sagittal (C) magnetic resonance arthrography (MR-A) sequences and the correspondingly reconstructed plane in FPCT-A images (B and D) of the same artificial defect. The marked depth measurements show larger values for the FPCT-A images.
Fig 5. Fusion image of an osteochondral defect and the cast.

Fusion mode images of an anteromedial defect in flat panel computed tomography arthrography (pink background) and the corresponding cast (brighter layer) in two different planes (A and B). Osteochondral defect and cast sizes can be directly compared. There is a visible overlay due to a slightly larger cast, explaining the mean differences of the measurements. The cross mark represents a 1 cm scale for both layers.
Discussion
The main results of the current experimental study are: 1) The diagnostic capabilities are similarly excellent with both high resolution flat-panel computed tomography arthrography (FPCT-A) and magnetic resonance arthrography (MR-A). 100% of the lesions were detected with either technique. 2) Although there is very good cross-correlation between both techniques, FPCT-A was slightly more accurate than MR-A, which was most significant when measuring lesion depth. Those results are in line with previous studies proposing FPCT-A enables accurate evaluation of articular cartilage thickness and detection of cartilage defects [22–24]. This has even been demonstrated for smaller lesions compared to the defects in our study. Similarly, first in vivo results focusing on the wrist FPCT-A report an advantage over MR-A in regards to the visualization of the intrinsic ligaments and the cartilage [19].
FPCT-A versus MR-A in the clinical context
For some time, MRI has been considered the standard procedure for the preoperative assessment of cartilage damage and localization to plan surgical access and the surgical type of procedure [27]. Recently, FPCT, a CT scanner model including volumetric imaging with ultra-high spatial resolution due to digital flat-panel detectors, was introduced [28]. This study suggests even a slightly superior performance of FPCT-A compared to MR-A regarding a precise osteochondral defect size measurement. We attribute the observed better correlations to FPCT´s higher spatial resolution of up to 0.11–0.12 mm in contrast to a 0.6 mm voxel size of the isotropic 3D MR sequence, the best possible resolution with the scanner. Thus, when interpreting the measurement differences, it is important to bear in mind the different resolutions associated with the different imaging modalities. An additional advantage of FPCT-A is the possibility of a free image plane adjustment. Although MR-A offers the advantage of completely avoiding any radiation exposure to the patient, The radiation dose is low for FPCT-A and comparable to a multi detector CT. Nevertheless, in the clinical setting, this needs consideration.
Technical nuances and discrepancies
The defect length, width and depth were measured smaller in FPCT-A compared to the cast specimens. These differences were not very pronounced, as neither value did exceed 1.1 mm. Nonetheless, as the results were significant, we believe the differences are accounted for by technical limitations of the cast preparation. Especially defect surface measurements could have been somewhat overestimated for the cast measurements due to the narrow overhanging lines at the defect margins. Differences in the depth measurements could be additionally explained also by difficulties with the precise determination of the formal chondral layer, which is especially challenging at the talar edges. Some variation of the measurements between the different modalities can possibly be attributed to a range of artifacts caused by intra-articular air collections. This intra-articular air impaired the MR-A images to a greater extent than the FPCT-A measurements due to susceptibility artifacts.
An important result was FPCT-A measured lesion depth better (deeper) than MR-A. Vice versa, MR-A appeared to underestimate defect depth. This result is in line with previous studies showing a general trend of suboptimal defect size estimation by MR-A [29–31]. Campbell et al. demonstrated that preoperative knee MRIs underestimated the size of articular cartilage defects compared to arthroscopic results [30]. In some cases, MRI can even fail to detect osteochondral lesions altogether [32].
Still, there was overall good correlation between FPCT-A and MR-A measurements, as has been recently published in a recent study by Deng et al., [33]. Both, MRI and CT were reliable in the assessment of the size of subchondral cysts of talar defects and the MRI classification was well-correlated with the CT classification [33].
Limitations
As the lesions evaluated in our study are artificial (surgically created), they do not fully resemble natural posttraumatic osteochondral lesions with chondral fissures, cartilage de-laminations or subtle subchondral defects. Accordingly, caution is needed when applying results of this study to the clinical practice.
As there were no ankles without osteochondral defects, there is no negative control group in this study. The radiologists knew to expect 4 lesions with every specimen. Thus, conclusions regarding sensitivity and specificity of the modalities are limited. Still, 100% of the lesions could be detected reliably with either method.
FPCT-A is a relatively new up and coming technique with only few reports discussing both in vivo and ex vivo defects [10, 34]. The cadaver specimens could be exposed to varying degrees of tissue de-composition, affecting the microstructure of tendons and cartilage. This could influence measurements using MR-A. Further, bone marrow edema is non-existent in the atraumatic cadaveric specimens, which would influence measurements in vivo.
A potential confounder may be the reference standard itself. Although the surface of the cast was essentially in line with the residual cartilage, the cast height (standard of the lesion depth) was measured systematically larger compared to both imaging modalities quantifying the lesion depth initially. This finding is reflected in the Bland-Altman plots. Nevertheless, this measurement approach is as close as possible to the real defect and cartilage thickness.
Conclusions
FPCT-A and MR-A both demonstrate excellent diagnostic performance in detecting and quantifying osteochondral defects in cadaveric ankle joints (ex vivo). FPCT-A appears to quantify defect dimensions slightly better, specifically the defect depth.
Acknowledgments
The authors would like to thank Prof. Erik Hauck for his help with manuscript revision and editing.
Abbreviations and acronyms
- CI
confidence interval
- FPCT-A
flat-panel computed tomography arthrography
- FS
fat suppression
- ICC
intraclass correlation coefficient
- MDCT
multidetector computed tomography
- MRI
magnetic resonance imaging
- MR-A
magnetic resonance arthrography
- PACS
picture archiving and communication system
- PDw
proton density weighted
- SD
standard deviation
- SPACE
sampling perfection with application optimized contrasts using different flip angle evolution
- T1w / T2w
T1 weighted / T2 weighted
- TSE
turbo spin echo
Data Availability
All relevant data are within the manuscript.
Funding Statement
This study has received funding by the Alwin Jäger Stiftung, Cranachstr. 56, 63739 Aschaffenburg, Germany. The grand recipient was the author Prof. Christoph Becher (CB). The grant amounts to 12.000€. http://alwinjaegerstiftung.de/ The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Martijn HA, Lambers KTA, Dahmen J, Stufkens SAS, Kerkhoffs GMMJ High incidence of (osteo)chondral lesions in ankle fractures. Knee Surgery, Sport Traumatol Arthrosc. 2021;29(5):1523–1534. doi: 10.1007/s00167-020-06187-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Murawski CD, Kennedy JG. Operative treatment of osteochondral lesions of the talus. J Bone Jt Surg—Ser A. 2013;95(11):1045–1054. doi: 10.2106/JBJS.L.00773 [DOI] [PubMed] [Google Scholar]
- 3.Hannon CP, Smyth NA, Murawski CD, et al. Osteochondral lesions of the talus: Aspects of current management. Bone Jt J. 2014;96 B(2):164–171. doi: 10.1302/0301-620X.96B2.31637 [DOI] [PubMed] [Google Scholar]
- 4.Raikin SM, Elias I, Zoga AC, Morrison WB, Besser MP, Schweitzer ME. Osteochondral lesions of the talus: Localization and morphologic data from 424 patients using a novel anatomical grid scheme. Foot Ankle Int. 2007;28(2):154–161. doi: 10.3113/FAI.2007.0154 [DOI] [PubMed] [Google Scholar]
- 5.Becher C, Thermann H. Results of microfracture in the treatment of articular cartilage defects of the talus. Foot Ankle Int. 2005;26(8):583–589. doi: 10.1177/107110070502600801 [DOI] [PubMed] [Google Scholar]
- 6.Giannini S, Buda R, Faldini C, et al. Surgical treatment of osteochondral lesions of the talus in young active patients. J Bone Jt Surg—Ser A. 2005;87(12 II SUPPL.):28–41. doi: 10.2106/JBJS.E.00516 [DOI] [PubMed] [Google Scholar]
- 7.Hurley ET, Stewart SK, Kennedy JG, Strauss EJ, Calder J, Ramasamy A. Current management strategies for osteochondral lesions of the talus. Bone Joint J. 2021;103-B(2):207–212. doi: 10.11302/0301-620X.103 [DOI] [PubMed] [Google Scholar]
- 8.Ramponi L, Yasui Y, Murawski CD, et al. Lesion Size Is a Predictor of Clinical Outcomes after Bone Marrow Stimulation for Osteochondral Lesions of the Talus: A Systematic Review. Am J Sports Med. 2017;45(7):1698–1705. doi: 10.1177/0363546516668292 [DOI] [PubMed] [Google Scholar]
- 9.Van Bergen CJA, Tuijthof GJM, Maas M, Sierevelt IN, Van Dijk CN. Arthroscopic accessibility of the talus quantified by computed tomography simulation. Am J Sports Med. 2012;40(10):2318–2324. doi: 10.1177/0363546512455403 [DOI] [PubMed] [Google Scholar]
- 10.El-Khoury GY, Alliman KJ, Lundberg HJ, Rudert MJ, Brown TD, Saltzman CL. Cartilage thickness in cadaveric ankles: measurement with double-contrast multi-detector row CT arthrography versus MR imaging. Radiology. 2004;233(3):768–773. doi: 10.1148/radiol.2333031921 [DOI] [PubMed] [Google Scholar]
- 11.Cerezal L, Llopis E, Canga A, Rolón A. MR Arthrography of the Ankle: Indications and Technique. Radiol Clin North Am. 2008;46(6):973–994. doi: 10.1016/j.rcl.2008.09.002 [DOI] [PubMed] [Google Scholar]
- 12.Hembree WC, Wittstein JR, Vinson EN, et al. Magnetic resonance imaging features of osteochondral lesions of the talus. Foot Ankle Int. 2012;33(7):591–597. doi: 10.3113/FAI.2012.0591 [DOI] [PubMed] [Google Scholar]
- 13.van Bergen CJA, Gerards RM, Opdam KTM, Terra MP, Kerkhoffs GMMJ. Diagnosing, planning and evaluating osteochondral ankle defects with imaging modalities. World J Orthop. 2015;6(11):944–953. doi: 10.5312/wjo.v6.i11.944 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Bauer JS, Barr C, Henning TD, et al. Magnetic resonance imaging of the ankle at 3.0 tesla and 1.5 tesla in human cadaver specimens with artificially created lesions of cartilage and ligaments. Invest Radiol. 2008;43(9):604–611. doi: 10.1097/RLI.0b013e31817e9ada [DOI] [PubMed] [Google Scholar]
- 15.Acid S, Corroller T Le, Aswad R, Pauly V, Champsaur P. Preoperative imaging of anterior shoulder instability: Diagnostic effectiveness of MDCT arthrography and comparison with MR arthrography and arthroscopy. Am J Roentgenol. 2012;198(3):661–667. doi: 10.2214/AJR.11.7251 [DOI] [PubMed] [Google Scholar]
- 16.Kyriakou Y, Struffert T, Dörfler A, Kalender WA. Grundlagen der Flachdetektor-CT (FD-CT). Radiologe. 2009;49(9):811–819. doi: 10.1007/s00117-009-1860-9 [DOI] [PubMed] [Google Scholar]
- 17.Chemouni D, Champsaur P, Guenoun D, Desrousseaux J, Pauly V, Le Corroller T. Diagnostic performance of flat-panel CT arthrography for cartilage defect detection in the ankle joint: Comparison With MDCT arthrography with gross anatomy as the reference standard. Am J Roentgenol. 2014;203(5):1069–1074. doi: 10.2214/AJR.13.12163 [DOI] [PubMed] [Google Scholar]
- 18.Gupta R, Grasruck M, Suess C, et al. Ultra-high resolution flat-panel volume CT: Fundamental principles, design architecture, and system characterization. Eur Radiol. 2006;16(6):1191–1205. doi: 10.1007/s00330-006-0156-y [DOI] [PubMed] [Google Scholar]
- 19.Sonnow L, Koennecker S, Luketina R, et al. High-resolution flat panel CT versus 3-T MR arthrography of the wrist: initial results in vivo. Eur Radiol. 2019;29(6):3233–3240. doi: 10.1007/s00330-018-5901-5 [DOI] [PubMed] [Google Scholar]
- 20.Werncke T, Sonnow L, Meyer BC, et al. Ultra-high resolution C-Arm CT arthrography of the wrist: Radiation dose and image quality compared to conventional multidetector computed tomography. Eur J Radiol. 2017;89:191–199. doi: 10.1016/j.ejrad.2017.02.009 [DOI] [PubMed] [Google Scholar]
- 21.Verhagen RAW, Maas M, Dijkgraaf MGW, Tol JL, Krips R, van Dijk CN. Prospective study on diagnostic strategies in osteochondral lesions of the talus. Is MRI superior to helical CT? J Bone Jt Surg—Ser B. 2005;87(1):41–46. doi: 10.1302/0301-620X.87B1.14702 [DOI] [PubMed] [Google Scholar]
- 22.Guggenberger R, Fischer MA, Hodler J, Pfammatter T, Andreisek G. Flat-Panel CT Arthrography: feasibility study and comparison to multidetector CT arthrography. Invest Radiol. 2012;47(5):312–318. doi: 10.1097/RLI.0b013e3182419082 [DOI] [PubMed] [Google Scholar]
- 23.Guggenberger R, Winklhofer S, Spiczak J V., Andreisek G, Alkadhi H. In vitro high-resolution flat-panel computed tomographic arthrography for artificial cartilage defect detection: Comparison With multidetector computed tomography. Invest Radiol. 2013;48(8):614–621. doi: 10.1097/RLI.0b013e318289fa78 [DOI] [PubMed] [Google Scholar]
- 24.Pagliano S, Chemouni D, Guggenberger R, et al. Correction to: Flat-panel CT arthrography for cartilage defect detection in the ankle joint: first results in vivo. Skeletal Radiol. 2020;49(8):1267. doi: 10.1007/s00256-020-03398-9 [DOI] [PubMed] [Google Scholar]
- 25.Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Cicchetti D V. Interreliability Standards in Psychological Evaluations. Psychol Assess. 1994;6(4):284–290. [Google Scholar]
- 27.Linklater JM. Imaging of talar dome chondral and osteochondral lesions. Top Magn Reson Imaging. 2010;21(1):3–13. doi: 10.1097/RMR.0b013e31820ef1c2 [DOI] [PubMed] [Google Scholar]
- 28.Reichardt B, Sarwar A, Bartling SH, et al. Musculoskeletal applications of flat-panel volume CT. Skeletal Radiol. 2008;37(12):1069–1076. doi: 10.1007/s00256-008-0473-0 [DOI] [PubMed] [Google Scholar]
- 29.Jung HG, Carag JA V., Park JY, Kim TH, Moon SG. Role of arthroscopic microfracture for cystic type osteochondral lesions of the talus with radiographic enhanced MRI support. Knee Surgery, Sport Traumatol Arthrosc. 2011;19(5):858–862. doi: 10.1007/s00167-011-1411-6 [DOI] [PubMed] [Google Scholar]
- 30.Campbell AB, Quatman CE, Flanigan DC, et al. Is Magnetic Resonance Imaging Assessment of the Size of Articular Cartilage Defects Accurate. J Knee Surg. 2013;27(1):67–76. [DOI] [PubMed] [Google Scholar]
- 31.Campbell AB, Knopp M V., Kolovich GP, et al. Preoperative MRI underestimates articular cartilage defect size compared with findings at arthroscopic knee surgery. Am J Sports Med. 2013;41(3):590–595. doi: 10.1177/0363546512472044 [DOI] [PubMed] [Google Scholar]
- 32.Tamam C, Tamam MO, Yildirim D, Mulazimoglu M. Diagnostic value of single-photon emission computed tomography combined with computed tomography in relation to MRI on osteochondral lesions of the talus. Nucl Med Commun. 2015;36(8):808–814. doi: 10.1097/MNM.0000000000000323 [DOI] [PubMed] [Google Scholar]
- 33.Deng E, Gao L, Shi W, et al. Both Magnetic Resonance Imaging and Computed Tomography Are Reliable and Valid in Evaluating Cystic Osteochondral Lesions of the Talus. Orthop J Sport Med. 2020;8(9):1–7. doi: 10.1177/2325967120946697 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Guggenberger R, Morsbach F, Alkadhi H, et al. C-arm flat-panel CT arthrography of the wrist and elbow: First experiences in human cadavers. Skeletal Radiol. 2013;42(3):419–429. doi: 10.1007/s00256-012-1501-7 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data are within the manuscript.