Skip to main content
. 2021 Aug 12;11:16430. doi: 10.1038/s41598-021-95799-6

Table 1.

RT-LAMP sensitivity across different ranges of Ct values.

RT-PCR (Ct value range)** RT-LAMP
True positives False negatives % Sensitivity (95% CI)*
[0–40] 84 42 67.2 (58.56–74.81)
[0–32] 68 0 100 (94.65–100)
[32–35] 17 20 45.95 (31.04–61.62)
[35–40] 11 27 28.95 (17–44.76)

*Wilson’s binominal confidence interval.

**Ct values rounded to the nearest unit.