Skip to main content
. 2020 Oct 7;45(7):657–665. doi: 10.1093/jat/bkaa137

Table IV.

PHMA ClinChek Results (µg/L) by Reference Material Level and Method

Ref. Mat. Observed mean SD CV (%) N % Error Established mean value Control range
Method
ClinChek Level I
A 4.41 0.728 16.5 5 –8.7 4.83 3.86–5.79
B 5.37 0.438 8.17 3 11.2
C 4.25 0.256 6.01 5 –12.0
D 5.11 0.272 5.32 5 5.8
E 4.44 0.399 8.99 5 –8.1
ClinChek Level II
A 38.5 5.87 15.3 5 –12.5 44.0 35.2–52.8
B 48.5 3.58 7.37 3 10.2
C 38.0 1.53 4.03 5 –13.6
D 46.7 3.53 7.56 5 6.1
E 41.7 2.94 7.05 5 –5.2

% Error was calculated as [(observed mean − established mean value)/established mean value] × 100.