Table V.
Unknown Urine Sample Results (µg/L) by Method (n = 50)
Method | LOD (µg/L) |
ULOQ (µg/L) |
n > LOD | Measured PHMA (µg/L) | |
---|---|---|---|---|---|
Minimum | Maximum | ||||
A | 0.600 | 100 | 40 | 0.607 | 9.03 |
B | 0.150 | 100 | 35 | 0.202 | 18.7 |
C | 0.080 | 120 | 42 | 0.094 | 26.2 |
D | 0.024 | 100 | 47 | 0.035 | 29.7 |
Ea | 0.200 | 50.0 | 45 | 0.201 | 16.6 |
Laboratory E uses a limit of quantitation (LOQ) rather than an LOD. For the purposes of this study, we used their LOQ as the method LOD. ULOQ, upper limit of quantitation.