. 2021 Jul 17;22(8):45–59. doi: 10.1002/acm2.13068

Table 1.

Linear accelerator acceptance tests and specifications. External references for tests and specifications are shown in brackets

Section Description Phantom Specification
A1 Safety
Inhibit systems
A2 Radiation shielding
Scattered radiation to the patient Mini‐phantom ave < 0.1% of in‐field dose (IEC 60601‐2‐1)
Bunker protection <1.0 mSv/year
A3 Coordinate systems and data integrity
Radiation beam and beam shaping (IEC61217)
MVI (IEC61217)
MRI (DICOM)
A4 MVI
Panel rigidity Ball bearing <0.3 mm
Rotational alignment Water phantom <0.2o
Pixel scale and isocenter Elekta pixel tool, MV alignment phantom
Image quality QC3, Las Vegas
A5 Mechanical and dosimetric alignment of gantry, focal spot and beam
Gantry tilt Spirit level <0.2o
Gantry rotation and readout Spirit level and spoke film <0.2o
Beam alignment Film, water phantom <1 mm and <0.2o
Isocenter locus Film <0.5 mm radius
A6 Mechanical and dosimetric alignment of MLC and jaws
MLC stripe test Film, MVI <0.5 mm (TG 142)
MLC transmission Film, water phantom <0.5% from baseline/TPS (TG142)
Jaw stripe test Film, MVI <1 mm (TG 142)
Gantry angle dependency Film, MVI <0.5 mm (TG 142)
A7 Table
Orthogonality and movement MVI <1° and <2 mm (TG 142)
A8 Laser — Elekta indicative sagittal laser Ball bearing <1 mm (TG 142)
A9 MR to MV alignment Elekta MR‐MV alignment phantom <0.3o (Elekta)
A10 Dosimetric system
Output stability — short and long term RW3 <2% (TG 142)
2nd linac monitor chamber RW3 <2% (TG 142)
Linearity RW3 <2% (TG 142)
Dose rate dependency RW3 <2% (TG 142)
Output gantry angle and cryostat dependency Mini‐phantom <1% from baseline (TG 142)
Profile constancy StarCheckMaxi <1% (TG 142)
Profile gantry angle dependency ICprofiler <1% (TG 142)
Internal monitoring and beam cutoff ICprofiler 3%
Reference dosimetry Water phantom <1% (NCS 18)
A11 Beam performance during MR imaging Water phantom <2%
A12 End‐to‐end test Alderson phantom

γ (5%/2mm),

10% threshold,

>90% pixels passed