Table 1.
Section | Description | Phantom | Specification |
---|---|---|---|
A1 | Safety | ||
Inhibit systems | |||
A2 | Radiation shielding | ||
Scattered radiation to the patient | Mini‐phantom | ave < 0.1% of in‐field dose (IEC 60601‐2‐1) | |
Bunker protection | <1.0 mSv/year | ||
A3 | Coordinate systems and data integrity | ||
Radiation beam and beam shaping (IEC61217) | |||
MVI (IEC61217) | |||
MRI (DICOM) | |||
A4 | MVI | ||
Panel rigidity | Ball bearing | <0.3 mm | |
Rotational alignment | Water phantom | <0.2o | |
Pixel scale and isocenter | Elekta pixel tool, MV alignment phantom | ||
Image quality | QC3, Las Vegas | ||
A5 | Mechanical and dosimetric alignment of gantry, focal spot and beam | ||
Gantry tilt | Spirit level | <0.2o | |
Gantry rotation and readout | Spirit level and spoke film | <0.2o | |
Beam alignment | Film, water phantom | <1 mm and <0.2o | |
Isocenter locus | Film | <0.5 mm radius | |
A6 | Mechanical and dosimetric alignment of MLC and jaws | ||
MLC stripe test | Film, MVI | <0.5 mm (TG 142) | |
MLC transmission | Film, water phantom | <0.5% from baseline/TPS (TG142) | |
Jaw stripe test | Film, MVI | <1 mm (TG 142) | |
Gantry angle dependency | Film, MVI | <0.5 mm (TG 142) | |
A7 | Table | ||
Orthogonality and movement | MVI | <1° and <2 mm (TG 142) | |
A8 | Laser — Elekta indicative sagittal laser | Ball bearing | <1 mm (TG 142) |
A9 | MR to MV alignment | Elekta MR‐MV alignment phantom | <0.3o (Elekta) |
A10 | Dosimetric system | ||
Output stability — short and long term | RW3 | <2% (TG 142) | |
2nd linac monitor chamber | RW3 | <2% (TG 142) | |
Linearity | RW3 | <2% (TG 142) | |
Dose rate dependency | RW3 | <2% (TG 142) | |
Output gantry angle and cryostat dependency | Mini‐phantom | <1% from baseline (TG 142) | |
Profile constancy | StarCheckMaxi | <1% (TG 142) | |
Profile gantry angle dependency | ICprofiler | <1% (TG 142) | |
Internal monitoring and beam cutoff | ICprofiler | 3% | |
Reference dosimetry | Water phantom | <1% (NCS 18) | |
A11 | Beam performance during MR imaging | Water phantom | <2% |
A12 | End‐to‐end test | Alderson phantom |
γ (5%/2mm), 10% threshold, >90% pixels passed |