Skip to main content
. 2021 Jul 11;22(8):148–155. doi: 10.1002/acm2.13339

TABLE 1.

iView panel to isocenter plane measured distances in millimeters

Machine iView panel location for panel to isocenter measurement (mm)
Gantry‐A Target‐A Gantry‐B Target‐B Center
A 588 593 589 595 592
B 589 594 590 595 593
C 593 598 593 598 595
D 590 595 589 596 593
E 580 581 581 580 579
F 585 586 586 587 586
G 583 586 584 587 585
H 589 594 590 595 593
I 590 594 590 594 591
J 598 601 598 601 600
Mean 588 592 589 593 591
St Dev 5 6 5 6 6
Range (Max–Min) 18 20 18 21 21