TABLE 1.
iView panel to isocenter plane measured distances in millimeters
| Machine | iView panel location for panel to isocenter measurement (mm) | ||||
|---|---|---|---|---|---|
| Gantry‐A | Target‐A | Gantry‐B | Target‐B | Center | |
| A | 588 | 593 | 589 | 595 | 592 |
| B | 589 | 594 | 590 | 595 | 593 |
| C | 593 | 598 | 593 | 598 | 595 |
| D | 590 | 595 | 589 | 596 | 593 |
| E | 580 | 581 | 581 | 580 | 579 |
| F | 585 | 586 | 586 | 587 | 586 |
| G | 583 | 586 | 584 | 587 | 585 |
| H | 589 | 594 | 590 | 595 | 593 |
| I | 590 | 594 | 590 | 594 | 591 |
| J | 598 | 601 | 598 | 601 | 600 |
| Mean | 588 | 592 | 589 | 593 | 591 |
| St Dev | 5 | 6 | 5 | 6 | 6 |
| Range (Max–Min) | 18 | 20 | 18 | 21 | 21 |