Fig. 6.

Fig. 6.

(A) Measured optical attenuation coefficient error versus tissue surface position error. (B) Sensitivity or specificity reduces as the quantification error of the optical attenuation coefficient increases. An ∼ 0.5 mm of surface position error induced an optical attenuation coefficient error of ∼ 0.1 mm-1 and reduced sensitivity or specificity by ∼ 1% (as indicated by the red dashed lines).