Summary of requirements on key control and monitoring parameters.

Parameter Range Tolerance Alarmsa Referenceb
Inspiratory-to-expiratory ratio 1:1-1:4 c Line 56
Respiratory rate (brpm) 10-30 ±2 Line 57
Tidal volume (mL) 350-450d (250-600)e ±4%f (± 15%)f ±20% Line 59
Inspiratory pressure (cmH2Og) <40 ±4% High, low, continuous Line 256
FiO2h (%) External controli External control External control External control

aAlarm conditions for these parameters are generated in compliance with ISO 80601-2-12 [11], clause 12.

bAAMI CR503:2020 [10] line number.

cNot specified.

dThe required tidal volume range.

eThe recommended tidal volume range.

fThe tolerance for tidal volume is reported as bias linearity.

g1 cmH2O is equivalent to 98.07 Pa.

hFiO2: fraction of inspired oxygen.

iControl of fraction of inspired oxygen is established by connection to an external oxygen supply.