Table 1.
Parameter | Range | Tolerance | Alarmsa | Referenceb |
Inspiratory-to-expiratory ratio | 1:1-1:4 | —c | — | Line 56 |
Respiratory rate (brpm) | 10-30 | ±2 | — | Line 57 |
Tidal volume (mL) | 350-450d (250-600)e | ±4%f (± 15%)f | ±20% | Line 59 |
Inspiratory pressure (cmH2Og) | <40 | ±4% | High, low, continuous | Line 256 |
FiO2h (%) | External controli | External control | External control | External control |
aAlarm conditions for these parameters are generated in compliance with ISO 80601-2-12 [11], clause 12.
bAAMI CR503:2020 [10] line number.
cNot specified.
dThe required tidal volume range.
eThe recommended tidal volume range.
fThe tolerance for tidal volume is reported as bias linearity.
g1 cmH2O is equivalent to 98.07 Pa.
hFiO2: fraction of inspired oxygen.
iControl of fraction of inspired oxygen is established by connection to an external oxygen supply.