Table 4.
Parameter | Priority level | Description |
Electrical power lost | High | The 12 V electrical power is lost. The ASVa cannot run in this state, but the alarms will be powered by the standby battery. |
Maximum pressure | High | The patient-circuit pressure has exceeded the maximum pressure parameter. This indicates there may be a problem with the PIP valve, the hoses, the endotracheal tube, or the patient. The ASV continues running, but stops the compression of the bag at maximum pressure on each cycle. |
Pressure low | High | The patient-circuit pressure is low. Possible reasons are the ASV has stopped due to loss of the pneumatic system or a disconnected patient hose. |
Volume low | High or medium | In PC-ACb mode, this is a high alarm when the volume drops below 250 mL. In VC-ACc mode this is a medium alarm when the volume drops below 80% of maximum volume. The expected volume has not been met, possibly due to disconnected or kinked hose. |
Volume high | Medium | In VC-AC mode, this is a medium alarm when the pressure exceeds 120% of maximum volume. |
9 V battery low | Medium | The 9 V standby battery must be replaced to prevent failure of the Electrical Power Lost high-priority alarm. However, in the absence of this 9 V battery’s power, the ASV continues running normally. |
aASV: acute-shortage ventilator.
bPC-AC: pressure-controlled assist control.
cVC-AC: volume-controlled assist control.