Table 11.

Reagent real time stability

Month Lot Reagent blank, A420nma Sensitivity, mA420nm·L/mgb Accuracy, ratioc
0 001 0.055 42.4 1.05
0 002 0.065 45.1 1.07
0 003 0.065 44.7 1.06
7 001 0.072 42.0 1.02
10 001 0.070 40.9 1.03
10 002 0.077 42.6 1.04
19 001 0.073 41.1 1.01
27 002 0.068 39.6 1.06
32 001 0.061 38.4 1.03
40 001 0.071 31.4 1.05
a

absorbance of the reagent blank at 420 nm.

b

Ratio of the absorbance (in milli-absorbance) obtained with the calibrator 3 by and the concentration of the same calibrator (10 mg/L).

c

Ratio of the concentration (mg/L) obtained in the control of the reagent and the and the concentration (mg/L) of the reference material.