TABLE 1.

Comparison of provider-collected NP swabs to patient-collected MT swabs and patient-collected saliva specimens for detection of SARS-CoV-2 RNA by real-time PCR

Sample type and SARS-CoV-2 result No. of NP swabs with a SARS-CoV-2 result of:
Detected Indeterminate Undetected
MT swab
    Detected 29 0 3a
    Indeterminate 0 0 0
    Undetected 3b 1c 245
Saliva
    Detected 28 0 2d
    Indeterminate 2e 0 2f
    Undetected 2g 1h 244
a

One of these 3 samples repeated as detected (i.e., positive) in both the NP and MT swabs. The remaining two samples repeated as undetected (i.e., negative) in both the NP and MT swabs. Both MT swabs had an initial Cp value of 35.0, suggesting that a low level of SARS-CoV-2 RNA was present in these specimens.

b

One of these 3 samples repeated as detected in both the NP and MT swabs. The remaining two samples repeated as detected in the NP swab and undetected in the MT swab. Both NP swabs had an initial Cp value of 35.0, suggesting that a low level of SARS-CoV-2 RNA was present in these specimens.

c

This MT swab sample repeated as undetected, while the NP swab was indeterminate upon repeat testing.

d

One of these samples repeated as detected in both the NP swab and saliva specimen, while the second sample repeated as undetected in the NP swab and indeterminate in the saliva specimen. The saliva specimen had an initial Cp value of 35.0, suggesting that a low level of SARS-CoV-2 RNA was present in this specimen.

e

One of these samples repeated as detected in both the NP swab and saliva specimen, while the second sample repeated as detected in the NP swab and indeterminate in the saliva specimen.

f

One sample repeated as undetected in both the NP swab and saliva specimen. The second sample repeated as undetected in the NP swab and detected in the saliva specimen.

g

Both saliva specimens repeated as undetected, while both NP swabs were detected upon repeat testing. Both NP swabs had an initial Cp value of 35.0, suggesting that a low level of SARS-CoV-2 RNA was present in these specimens.

h

The NP swab repeated as indeterminate, while the saliva specimen repeated as undetected upon repeat testing.