TABLE 3.
Metric | No. (%) | Pool N1 or N2 CT value of <30, no. (%) |
---|---|---|
Positive pools with available reflex test data | 124 | 86/124 (69.4) |
Pools with complete BinaxNOW deconvolution (all individuals tested by BinaxNOW) | 114/124 (91.9) | 80/114 (70.2) |
True-positive BinaxNOWa | 89/114 (78.1) | 74/89 (83.1) |
False-negative BinaxNOWb | 5/114 (4.4) | 1/5 (20.0) |
Likely false-negative BinaxNOWc | 4/114 (3.5) | 1/4 (25.0) |
Pool PCR result unconfirmedd | 12/114 (10.5) | 2/12 (16.7) |
Unknowne | 4/114 (3.5) | 2/4 (50.0) |
Individual in pool tested positive by BinaxNOW (in 88/89 cases, only one positive BinaxNOW result was observed; in 1/89, two were observed [the two individuals were from the same family]).
All BinaxNOW results were negative, but an individual in the pool tested positive by reflex PCR.
All BinaxNOW results were negative, but an individual in the pool had an inconclusive reflex PCR result (only 1 of 2 PCR targets positive).
All BinaxNOW and additional reflex testing results (repeat BinaxNOW testing (n = 4) or PCR testing (n = 8) on all individuals in the pool) were negative, making it impossible to confirm the positive pool PCR result and leaving open the possibility that the PCR result was a false positive.
All BinaxNOW results were negative, but additional reflex PCR results were not available.