. 2021 Aug 19;8:61. doi: 10.1186/s40658-021-00407-1

Table 1.

CFs for 123I with the LEHR and ME collimator and 131I with the HE collimator, in cps/MBq, for the four CF

source geometries

VOI diameter 123I LEHR (cps/MBq) 123I ME (cps/MBq) 131I HE (cps/MBq)
1 CFcylinder 89.3 119.0 27.7
2 CFcylinder 88.4 115.6 26.8
CFwater 84.4 118.2 27.6
CFair 82.4 116.4 27.2
CFpoint 84.1 115.5 27.8
Average ± SD 84.8 ± 2.5 116.4 ± 1.3 27.4 ± 0.4
Relative SD (%) 2.9 1.1 1.5
3 CFcylinder 100.4 118.9 28.6
CFwater 102.2 120.2 29.1
CFair 98.4 117.7 29.4
CFpoint 102.4 118.7 31.7
Average ± SD 100.9 ± 1.9 118.9 ± 1.0 29.7 ± 1.4
Relative SD (%) 1.9 0.8 4.7

VOI diameters were defined as (1) less than the physical source diameter, (2) equal to the physical source diameter plus a 3-cm margin and (3) equal to the physical size of the detector