Table 4.

Summary of the in vivo biocorrosion and biocompatibility properties of biodegradable Zn-based BMs for cardiovascular applications.

Material Implant shape Animal mode Implantation site Period time In vivo corrosion rate (mm/year) Inflammatory response Thrombosis Intimal hyperplasia Ref.
Pure Zn (99.995%) Stent Adult Japanese rabbit Abdominal aorta 12 months NA [164]
Pure Zn Wire Rat Abdominal aorta 11 months NA NA NA _ [188]
Pure Zn (99.99+%) Wire Male young CD IGS rat Abdominal aorta 6 months NA + NA NA [171]
Pure Zn (99.99+%) Wire Sprague Dawley rat Abdominal aorta 6.5 months NA _ _ _ [189]
Zn-0.002Mg Wire Adult male Sprague-Dawley rat Abdominal aorta 11 months ~0.027 at 4.5 months and ~0.050 at 11 months + NA + [167]
Zn-0.005Mg Wire Adult male Sprague-Dawley rat Abdominal aorta 11 months ~0.023 at 4.5 months and ~0.039 at 11 months + NA + [167]
Zn-0.08Mg Wire Adult male Sprague-Dawley rat Abdominal aorta 11 months ~0.015 at 4.5 months and ~0.023 at 11 months + NA + [167]
Zn-0.8Cu Stent Female Shanghai white pig Left anterior descending artery, left circumflex artery, or right coronary artery 24 months ~0.016 at 3 months + + [165]
Zn–3Ag Stent Juvenile domestic swine Iliofemoral artery 6 months NA NA NA [166]
Zn-0.1Li Wire Sprague Dawley rat Abdominal aorta 12 months ~0.019 at 9 months and ~0.046 at 12 months + NA + [168]
Zn–4Li Wire Rat Abdominal aorta 11 months NA NA NA + [188]
Zn-0.02Mg-0.02Cu Stent New Zealand white rabbit Left carotid artery 12 months ~0.025 at 3–6 months and 0.040 ± 0.013 at 12 months NA NA + [183]

Note: +: Positive; -: Negative; NA: Not available.