Comparison of POC whole blood fingerstick with POC serum.

  Results of LDBIO Kit whole blood (30 µL) fingerstick VS LDBIO Kit (15 µL) of serum obtained by venipuncture
  POC
Whole blood test
POC
Serum test
Western Blot TOXO II test
Positive 108 112 112
Negative 236 237 237
  5 discordant (1 false positive and 4 false negatives) Total: 349
Performance Sensitivity: 96.4% [IC95 90.6–98.9%]
Specificity: 99.6% [IC95 97.3–99.9%]