Table 6. Percent agreement between the STANDARD G6PD Test and the spectrophotometric reference test using the manufacturer’s threshold values at 30% and 70% G6PD activity thresholds, on venous specimens.

Spectrophotometric reference test
Deficient Intermediate Normal Total
STANDARD G6PD Test Deficient 56 14 9 79
Intermediate 0 17 31 48
Normal 0 1 1,534 1,535
Total 56 32 1,574 1,662

G6PD, glucose-6-phosphate dehydrogenase.

Kappa: 0.73. Percent agreement between hemoglobin-normalized G6PD activity categorized results and the STANDARD Test was 96.7% (95% CI 95.7%–97.5%).