Table 7. Percent agreement between the STANDARD G6PD Test and the spectrophotometric reference test using the manufacturer’s threshold values at 30% and 70% G6PD activity thresholds, on capillary specimens.

Spectrophotometric reference test
Deficient Intermediate Normal Total
STANDARD G6PD Test Deficient 58 17 19 94
Intermediate 0 16 62 78
Normal 0 2 1,519 1,521
Total 58 35 1,600 1,693

G6PD, glucose-6-phosphate dehydrogenase.

Kappa: 0.60. Percent agreement between hemoglobin-normalized G6PD activity categorized results and the STANDARD Test was 94.1% (95% CI 92.9%–95.2%).