Skip to main content
. Author manuscript; available in PMC: 2022 Sep 10.
Published in final edited form as: J Pharm Biomed Anal. 2021 Jul 21;204:114274. doi: 10.1016/j.jpba.2021.114274

Table 4.

Summary of stability evaluation for JQ1 in mouse plasma and microdialysates (n=3).

Sample Matrix Stability Study Nominal Concentration (ng/mL) Mean ± S.D. Calculated Concentration (ng/mL) Precision (% R.S.D.) Accuracy (% R.E.) % Mean deviation
Mouse Plasma Process a 15 14.93 ± 1.10 7.34 −0.46 −3.80
800 686.39 ± 43.52 6.34 −14.20 −7.24
Bench-top RT b 15 16.27 ± 1.80 11.07 8.47 4.84
800 767.80 ± 29.98 3.90 −4.02 3.76
Bench-top 4C c 15 15.57 ± 0.98 6.30 3.80 0.33
800 767.45 ± 16.91 2.20 −4.07 3.71
Freeze-thaw d 15 14.40 ± 0.51 3.52 −3.98 −7.20
800 793.88 ± 23.31 2.94 −0.77 7.29
Long-term e 15 14.70 ± 0.69 4.69 −2.03 −5.31
800 788.56 ± 19.49 2.47 −1.43 6.57
Microdialysate Process a 1.5 1.45 ± 0.10 6.88 −3.09 0.46
400 388.38 ± 2.23 0.57 −2.91 1.39
Bench-top RT b 1.5 1.49 ± 0.06 3.89 −0.42 3.23
400 403.22 ± 4.61 1.14 0.81 5.26
Bench-top 4C c 1.5 1.53 ± 0.03 1.84 2.22 5.96
400 400.64 ± 13.48 3.37 0.16 4.59
Freeze-thaw d 1.5 1.62 ± 0.017 1.03 7.83 11.77
400 408.69 ± 6.31 1.54 2.17 6.69
Long-term f 1.5 1.63 ± 0.15 8.88 8.85 12.83
400 404.60 ± 6.46 1.60 1.15 5.62

%R.S.D.: relative standard deviation; %R.E.: relative error

a

Stability assessed after 48 h in auto sampler at 4°C

b

Short-term stability after 24h at room temperature

c

Short-term stability after 24h at 4°C

d

Stability evaluated after three freeze-thaw cycles

e

91 days long-term stability in mouse plasma at −80°C

f

21 days long-term stability in microdialysate at −80°C