. 2021 Jul 14;6(4):e01330-20. doi: 10.1128/mSphere.01330-20

TABLE 3.

Diagnostic accuracy in measles and rubella antibody detection comparing paired serum and dried blood spot specimensa

Parameter % diagnostic accuracy (95% CI)
Treating equivocal results as positive results
Excluding equivocal results
Measles Rubella Measles Rubella
Sensitivity 98.6 (97.2, 99.5) 98.4 (96.8, 99.3) 99.4 (98.2, 99.9) 99.0 (97.6, 99.7)
Specificity 90.0 (73.5, 97.9) 98.1 (89.7, 100) 100 (87.2, 100) 100 (93.0, 100)
PPV 99.4 (98.3, 99.9) 99.8 (98.9, 100) 100 (99.3, 100) 100 (99.2, 100)
NPV 79.4 (62.1, 91.3) 86.4 (75.0, 94.0) 90.0 (73.5, 97.9) 91.1 (80.4, 97.0)
% agreement 98.1 98.3 99.4 99.1
Kappa statistic 0.83 (0.73, 0.93) 0.91 (0.85, 0.97) 0.94 (0.88, 1.00) 0.95 (0.90, 0.99)
a

Analysis excludes devices with contamination, extremely low volume, or mislabeling. 95% CI, 95% exact confidence intervals; PPV, positive predictive value; NPV, negative predictive value.