. 2021 Aug 26;16(8):e0255172. doi: 10.1371/journal.pone.0255172

Table 4. The diagnostic efficacy of each BDG detection kit (N = 164).

Sensitivity (%) Specificity (%) PPV (%) NPV (%) LR+ LR-
ES 90.9 (10/11) 91.5 (140/153) 43.5 (10/23) 99.3 (140/141) 10.7 0.1
FA 63.6 (7/11) 91.5 (140/153) 35.0 (7/20) 97.2 (140/144) 7.5 0.4
MKII 100 (11/11) 90.8 (139/153) 44.0 (11/25) 100 (139/139) 10.9 0.0
Wako 100 (11/11) 96.1 (147/153) 64.7 (11/17) 100 (147/147) 25.5 0.0
Wako-Eu 100 (11/11) 96.1 (147/153) 64.7 (11/17) 100 (147/147) 25.5 0.0

BGD, β-D-glucan; PPV, Positive predictive value; NPV, Negative predictive value; LR+, Positive likelihood ratio; LR-, Negative likelihood ratio; ES: Fungitec G test ES “Nissui”; FA, Fungitell β-D-glucan assay kit; MKII, Fungitec G test MKII “Nissui”; Wako, β-Glucan test Wako; Wako-Eu, European version of β-Glucan test Wako.