Table 1.

A comparison of the on-chip ultramicroelectrode glucose sensor, presented here with other reagent-free miniaturized enzymatic glucose biosensors, constructed towards various biomedical sensing applications.

Electrode Material * Linear Range Operating Potential Enzyme Immobilization Sensitivity Real Sample Reference
CFM/Pt-GOx(0.5%)-GA 0–4 mM 0.7 V vs. Ag/AgCl (external RE) Covalent 0.697 nA mM−1 cm−2 Brain slices [48]
CF-PEDOT-GOx 0.5–15 mM −0.65 V vs. Ag/AgCl (external RE) Entrapment 8.5 μA mM−1 cm−2 - [49]
Pt/PEDGE/BSA-GOx 0–5 mM 0.85 V vs. Ag/AgCl (external RE) Covalent 20.72 pA mM−1 Intracellular [50]
Pt-B electrode 0.2–43.5 mM 0.7 V vs. Ag/AgCl (on-device) Free in solution - Serum samples [51]
Pt/PAA/CS/AuNPs/GOx 0.017–0.81 mM −0.4 V to 0.4 V to vs. Pt (on-device) Layer-by-layer assembly 61.4 μA mM−1 cm−2 Saliva [52]
Silver/Nafion/GA-GOx 0–2 mM 0.05 V vs. Ag/AgCl (external electrode) Covalent 1.73 μA mM−1 - [53]
Pt/GA-BSA-GOx/PU/Nafion 0–11 mM 0.5 V vs. IrOx (on-device) Covalent 1.51 nA mM−1 Intradermal [54]
GOx-β-cyclodextrin-o-PD/Pt-B/Au 2.5–15 mM 0.25 V vs. Pt (on-device) Entrapment 111.21 µA mM−1 cm−2 (Buffer), 15 µA mM−1 cm−2 (Serum) Fetal bovine serum This work

* Abbreviations: CFM: carbon fibre microelectrode, Pt: platinum, GOx: glucose oxidase, GA: glutaraldehyde, CF: carbon fibre, PEDOT: poly (3, 4-ethylenedioxythiophene), PEDGE: poly (ethylene glycol) diglycidyl ether, BSA: bovine serum albumin, PAA: poly (allylamine), CS: chitosan, AuNPs: gold nanoparticles, and PU: polyurethane.