Table 1.
Comparison of performance of fiber sensors with different configurations for micro-force measurements
| Sensor structure | Size of sensor | Force sensitivity (pm μN-1) | Reference |
|---|---|---|---|
| Photonic crystal fiber | 125 μm × 3 cm | 0.016 × 10−3 | 54 |
| Microfiber Bragg grating | 2.4 mm × 2.4 mm | 0.73 × 10−3 | 55 |
| FP cuboid cavity | 18 μm × 60 μm | 0.026 | 56 |
| Microfiber asymmetrical FP interferometer | 20 mm × 7.3 um | 0.221 | 57 |
| FP micro-cavity plugged by cantilever taper | 1.36 mm × 125 μm | 0.842 | 58 |
| Balloon-like interferometer | 24 mm × 14 μm | 24.9 | 15 |
| Clamped beam probe | 68 μm × 100 μm | 1510 | This work |