Table 3.

Statistical Parameters That Determine the Efficacy of Diagnosing Subclinical and Clinical Keratoconus Using α Maps

Sensitivity (%) Specificity (%) Accuracy (%) Precision (%) AUC Cutoff Value (a.u.)
Control vs. subclinical keratoconus 90.0 95.0 92.5 94.7 0.97 45.1
Control vs. keratoconus 100 100 100 100 1.0 47.9