Table 2.
The target depths and critical PRR for different types of PAI systems. PRR, pulse repetition rate; PAI, photoacoustic imaging; PACT, photoacoustic computed tomography; AR-PAM, acoustic-resolution photoacoustic microscopy; OR-PAM, optical-resolution photoacoustic microscopy; , speed of sound in soft tissue.
PACT | AR-PAM | OR-PAM | |
---|---|---|---|
Optical focus regime | Diffusive | Quasi-diffusive | Ballistic regime |
Effective imaging depth | ≤ 5 cm | ≤ 10 mm | ≤ 1 mm |
Critical PRR | 30.8 kHz | 154 kHz | 1.54 MHz |
(where 1540 m/s).