Skip to main content
. 2021 May 6;50(6):20210005. doi: 10.1259/dmfr.20210005

Table 3.

Sensitivity, specificity, false-positive (FP) value, false-negative (FN) value and area under the curve (AUC) for NIRR from the occlusal and trilateral views as well as for digital BWR with µCT as the reference test (lower and upper 0.95 CI in parentheses) for caries lesions

Sensitivity Specificity FP FN AUC
NIRR occlusal All teeth 0.29
(0.20–0.39)
0.84
(0.78–0.90)
0.16
(0.10–0.22)
0.71
(0.61–0.77)
0.57
(0.49–0.64)
Premolar 0.39
(0.24–0.55)
0.84
(0.68–1.01)
0.16
(-0.01–0.32)
0.61
(0.45–0.76)
0.62
(0.47–0.77)
Molar 0.22
(0.11–0.33)
0.84
(0.78–0.90)
0.16
(0.10–0.22)
0.78
(0.67–0.84)
0.53
(0.44–0.62)
NIRR trilateral All teeth 0.63
(0.53–0.73)
0.70
(0.62–0.77)
0.30
(0.23–0.38)
0.37
(0.27–0.45)
0.66
(0.59–0.73)
Premolar 0.66
(0.51–0.81)
0.53
(0.30–0.75)
0.47
(0.25–0.70)
0.34
(0.19–0.54)
0.59
(0.43–0.75)
Molar 0.61
(0.48–0.74)
0.72
(0.64–0.79)
0.28
(0.21–0.36)
0.39
(0.26–0.48)
0.67
(0.58–0.75)
BWR All teeth 0.27
(0.17–0.36)
1.00
(1.00–1.00)
0.00
(0.00–0.00)
0.73
(0.64–0.80)
0.63
(0.55–0.70)
Premolar 0.33
(0.18–0.49)
1.00
(1.00–1.00)
0.00
(0.00–0.00)
0.67
(0.51–0.81)
0.65
(0.50–0.80)
Molar 0.22
(0.11–0.33)
1.00
(1.00-.100)
0.00
(0.00–0.00)
0.78
(0.67–0.84)
0.60
(0.50–0.69)