Table 5.

Uncertainty budget for the massic activity (Am) of the S3 224Ra solution determined at NPL by 4π(LS)-γ DCC.

Component ui/%
Standard uncertainty of the weighted mean 0.032
Gravimetric 0.050
Repeatability 0.10
γ-gate selection 0.10
Extrapolation 0.20
Decay correction 0.072
228Th impurity 0.010
Background correction 0.0030
Predicted equilibrium ratio of decay progeny 0.015
Pulse pile-up (LS-channel) 0.015
Pulse pile-up (γ-channel) 0.0050
LS-channel dead time 0.050
γ-channel dead time 0.050
Accidental coincidence correction 0.010

Combined standard uncertainty, uc 0.27