Table 6.

Uncertainty budget for the activity of ampoule S3-A, determined with the Vinten 671 ionization chamber (VIC) at NIST during the bilateral comparison.

Component ui/%
Measurement uncertainty; estimated as the relative standard deviation of the mean of N = 500 current measurements 0.36
Standard uncertainty on KVIC 0.32
Background 0.24
228Th impurity 0.01
Decay correction 0.001
Solution composition 0.18

Combined standard uncertainty, uc 0.57