Table 5.

Root mean square error (RMSE) values at different temperature ranges, calculated as the difference between tympanic temperature and forehead surface skin measurements (N = 177) using four pieces of equipment.


Root mean square error (RMSE)
Temperature Range (oC) IR Digital Contact Probe (Forehead) Thermal Imaging Camera iButton® FLIR Laser (0.61 m)
0-5 1.31 0.84 7.77 7.99
5.1–10 1.30 0.97 8.59 6.89
10.1–15 1.08 0.81 7.32 6.19
15.1–20 0.94 0.87 6.91 5.27
20.1–25 0.66 0.85 6.86 4.37
25.1–45 0.67 0.90 6.36 3.63
Overall 1.04 0.86 7.31 5.72