Table 2. Acquisition protocol recommendations for spatial resolution optimization in UHRCT.

Target parameter Line pairs improvement TTF10% improvement Recommendation
Reconstruction Matrix 8 28.2% The highest possible in standard/high-dose protocols. If low-dose is required use standard matrix setting
Tube output 5 29.6% Work with standard to high dose. Use the highest mA reasonably achievable
Reconstruction algorithm 4 93.3% Use MBIR. If not available use HIR with the sharpest kernel possible
Lateral displacement 4 30.8% Pay close attention to patient centering
Slice thickness 3 17.3% The smallest possible if high reconstruction reformats are required*
Field-of-view 2 15.1% The smallest possible
Focal spot size 0 24% Use automatic scanner setting

*, No influence on the “in-plane” spatial resolution. MTF, Modulation transfer function; mA, Tube current; HIR, Hybrid iterative reconstruction; MBIR, model based iterative reconstruction; TTF, Task transfer function.