Table 2.
Automated algorithm atrial fibrillation status | Manual atrial fibrillation status | ||
Atrial fibrillation | No atrial fibrillation | Total | |
Atrial fibrillation | 24 | 0 | 24 |
No atrial fibrillation | 0 | 25 | 25 |
Total | 24 | 25 | 49b |
aThe sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were 100% (95% CI 86%-100%), 100% (95% CI 86%-100%), 100%, 100%, and 100% (95% CI 93%-100%), respectively.
bOne patient had noise in the waveform tracings that were unable to be run through the algorithm due to lack of 3 consecutive 2-minute segments of majority noise-free time.