Table 2.

Results of training set 1a for initial automated atrial fibrillation identification using signal and noise detection and R–R interval indices, from selected waveforms without premature beats.

Automated algorithm atrial fibrillation status Manual atrial fibrillation status
Atrial fibrillation No atrial fibrillation Total
Atrial fibrillation 24 0 24
No atrial fibrillation 0 25 25
Total 24 25 49b

aThe sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were 100% (95% CI 86%-100%), 100% (95% CI 86%-100%), 100%, 100%, and 100% (95% CI 93%-100%), respectively.

bOne patient had noise in the waveform tracings that were unable to be run through the algorithm due to lack of 3 consecutive 2-minute segments of majority noise-free time.