Table 6.
Results of the test set 2a for automated atrial fibrillation signal and noise detection and R–R interval indices, with added P-wave and premature beat detection, from unselected waveforms.
Automated algorithm atrial fibrillation status | Manual atrial fibrillation status | ||
Atrial fibrillation | No atrial fibrillation | Total | |
Atrial fibrillation | 23 | 1 | 24 |
No atrial fibrillation | 2 | 24 | 26 |
Total | 25 | 25 | 50 |
aThe sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were 92% (95% CI 74%-99%), 96% (95% CI 80%-100%), 96% (95% CI 77%-99%), 92% (95% CI 76%-98%), and 94% (95% CI 83%-99%), respectively.