Table 3.

Performance of commercial kits using treated saliva as a template for RT-qPCR.

Commercial kit Sample size, n Invalid specimens, n Concordant to reference, n Invalid rate, % Overall agreement, %
Maccura 11 0 11 0.00 100.00
Fosun 12 1 11 8.33 100.00
Ardent 11 0 10 0.00 90.91
SD Biosensor 13 1 11 7.69 91.67
Biosewoom 11 1 8 9.09 80.00
Fortitude 11 1 7 9.09 70.00
3S 10 0 4 0.00 40.00
Vazyme 10 4 2 40.00 33.33

Rates of the invalid and overall agreement were calculated as compared to results from the reference, Detection Kit for 2019 Novel Coronavirus (2019-nCoV) by Da An Gene Co.