Skip to main content
. 2021 Sep 1;144(1):011004. doi: 10.1115/1.4051845

Fig. 3.

Measured AFO stiffness (mean [range]) across a total of four sessions (two per operator). Sagittal plane stiffness across AFOs varied from 0.58 to 3.66 N·m/deg of dorsiflexion. Most AFOs produced an eversion moment in the frontal plane, while the two AFOs with medial struts produced an inversion moment.

Measured AFO stiffness (mean [range]) across a total of four sessions (two per operator). Sagittal plane stiffness across AFOs varied from 0.58 to 3.66 N·m/deg of dorsiflexion. Most AFOs produced an eversion moment in the frontal plane, while the two AFOs with medial struts produced an inversion moment.