Skip to main content
. 2021 Sep 6;8(1):e000939. doi: 10.1136/bmjresp-2021-000939

Table 1.

Patient and pulse oximeter characteristics

Number/statistic
Patient characteristics (n=35)
 Age, years 69 (61–73)*
 Female 14 (40%)
 Fitzpatrick IV–VI skintype 5 (14.3%)
 Body temperature, oC 37.2±0.9
 Heart rate, beats per minute 84 (68–101)*
 Systolic blood pressure (mm Hg) 125 (108–143)*
 Diastolic blood pressure (mm Hg) 57 (50–66)*
 Requiring vasopressor 16/35 (45.7%)
 Noradrenalin dosage (mg/h) 0.4 (0.3–1.60)*
 Requiring supplemental oxygen 35/35 (100%)
Pulse oximeter reading characteristics (n=2258)
 AFAC FS10D 232/234 (99.1%)
 AGPTEK FS10C 233/234 (99.6%)
 ANAPULSE ANP 100 202/234 (86.3%)
 Cocobear 226/234 (96.6%)
 Contec CMS50D1 228/234 (97.4%)
 HYLOGY MD-H37 232/234 (99.1%)
 Mommed YM101 229/234 (97.9%)
 PRCMISEMED F4 PRO 233/234 (99.6%)
 PULOX-PO-200 231/234 (98.7%)
 Zacurate Pro Series 500 DL 212/234 (90.6%)
 Total pulse oximeter readings 2258/2340 (96.5%)

Binary variables presented as number and percentage. Continuous variables were displayed as mean±SD when normally distributed, or median and 25th–75th percentile (*) when non-normally distributed.