TABLE 7.
Example initial brachytherapy plan check items for medical physicists and their designees
Recommended | Optional | |
---|---|---|
Planning | Rx matches plan | |
Source activity/air kerma strength (against decay table) | ||
Correct planning image (eg, correct date/time) | ||
Plan normalization | ||
Channel number assignment follow local convention | ||
Index length/offset | ||
Catheter orientation (tip vs. connector ends) | ||
Number of dwell positions and location of the first dwell point | ||
Step size | ||
Applicator name, size, and model | ||
Magnified catheter reconstruction in 3D view | ||
Plan optimization appropriate | ||
Dose distribution appropriate | ||
Calculation algorithm | ||
Secondary calculation | ||
RO‐EMR preparation | Rx and plan approved by physician | |
Plan approved by physicist |