TABLE 7.
Example initial brachytherapy plan check items for medical physicists and their designees
| Recommended | Optional | |
|---|---|---|
| Planning | Rx matches plan | |
| Source activity/air kerma strength (against decay table) | ||
| Correct planning image (eg, correct date/time) | ||
| Plan normalization | ||
| Channel number assignment follow local convention | ||
| Index length/offset | ||
| Catheter orientation (tip vs. connector ends) | ||
| Number of dwell positions and location of the first dwell point | ||
| Step size | ||
| Applicator name, size, and model | ||
| Magnified catheter reconstruction in 3D view | ||
| Plan optimization appropriate | ||
| Dose distribution appropriate | ||
| Calculation algorithm | ||
| Secondary calculation | ||
| RO‐EMR preparation | Rx and plan approved by physician | |
| Plan approved by physicist |