TABLE 9.
Recommended | Optional |
---|---|
Patient name, MRN, DOB | Enter “custom” for Bolus fields |
Patient photos and setup photos | DRR approved and associated |
Consent signed by physician | MUs/daily dose reasonable |
Rx signed by physician | Peer review signed by physician/physicist |
Plan laterality matched with Rx in RO‐EMR | Treatment couch inserted in plan |
Plan used the scheduled treatment unit | Intra‐treatment image method (eg, triggered imaging) in RX |
Isocenter shifts instruction clear and documented in setup note | Fiducial contoured |
Rx dose/fractions in RO‐EMR agree with plan | Check corrected images input to the third party image systems |
Rx Tx technique in RO‐EMR agrees with plan | |
Rx energy in RO‐EMR agrees with plan | Reference CT/scan sent to third party system |
Field parameters consistency between plan and RO‐EMR (for third party RO‐EMR) | ROI box drawn appropriately |
SSD parameters | Bolus field entered |
Treatment fields approved by physicist | |
Treatment plan deliverability check (loading reference CT and collision check for non‐coplanar beams) | Physics consult presence |
Patient setup (immobilization device setup, skin markers) info verified | Dose rate |
Special alerts/dose actions (position, bolus, in‐vivo measurement, etc.) | Blocks/compensator in Rx and setup note |
Bolus documented | |
In‐vivo measurement indicated | |
Respiratory/IGRT instructions | |
Third party IGRT data ready | |
Motion management parameters | |
Treatment schedule | |
Boost fields are scheduled/follow‐up | |
Tx appointment conflicts, concurrent chemo documented | |
Implanted device presence/document | |
Second MU check documented | |
Physics check completed/approved | |
IMRT/VMAT QA completed/approved | |
Pre‐RT preparations | |
For electron fields: | |
e‐applicator field parameters (energy cone, field size, gantry) | |
e‐block checked to the template | |
e‐block code verify (accessory code) |
Abbreviations: DIBH, deep inspiration breath hold; DRR, digitally reconstructed radiograph; EDW, enhanced dynamic wedge; IGRT, image‐guided radiotherapy; KVCBCT, Kilo‐voltage cone beam CT; MU, monitor unit; MVCBCT, Mega‐voltage cone beam CT; QCL, quality‐check‐list; RO‐EMR, Radiation oncology specific electron medical record system; ROIs, regions‐of‐interest; Rx, prescription; SID, source‐imager distance; SSD, source‐skin distance; Tx, treatment.