Table 5.
Diagnostic performance of QFR and residual QFR to assess significant CAD (pre-PCI FFR ≤ 0.80) and a suboptimal PCI outcome (post-PCI FFR < 0.90), respectively.
Diagnostic performance | 128 vessels |
---|---|
QFR to detect pre-PCI FFR ≤ 0.80 | |
Sensitivity | 100% (97–100%) |
Specificity | 27% (6–61%) |
Negative predictive value | 100% (not applicable) |
Positive predictive value | 94% (91–95%) |
Accuracy | 94% (88–97%) |
| |
Residual QFR to detect post-PCI FFR < 0.90 | |
Sensitivity | 44% (31–59%) |
Specificity | 96% (87–99%) |
Negative predictive value | 70% (65–75%) |
Positive predictive value | 89% (72–96%) |
Accuracy | 74% (66–82%) |
Values are presented as % (95% confidence interval). FFR, fractional flow reserve; PCI, percutaneous coronary intervention; QFR, quantitative flow ratio.