Table 7.
Summary of the in-vivo wear studies (All the values listed below are the mean values)
| Researcher | Test subjects | Results | Testing technique |
|---|---|---|---|
| Geringer [58] | 12 explants of DMC | Convex: 36.7±29.1 mm3/year | In-vivo test (CMM) |
| Concave: 22.1±39.2 mm3/year | |||
| Total wear rate: 53.9±50.3 mm3/year | |||
| Imbert [22] | 8 explants of DMC | Convex: 29.72 mm3/year | In-vivo test (CMM) |
| Concave: 15.96 mm3/year | |||
| Boyer [23] | 35 explants of DMC | Convex: 25 mm3/year | In-vivo test (Light Surface Scanner) |
| Concave: 12 mm3/year | |||
| Adam [24] | 40 retrieved DM liners | convex: 28.9 mm3/year | In-vivo test (Surface analysis) |
| concave: 25.5 mm3/year | |||
| total wear rate: 54.3 mm3 year | |||
| Deckard [25] | 34 implants of DMC | Mean volumetric wear: | Martell methodology at regular postoperative intervals |
| 783 mm3/year at 1 year | |||
| 555 mm3/year at 2 years | |||
| 104 mm3/year at 5 years | |||
| Leonard [26] | 129 retrieved DM liners | The lever-out force for same manufacturer: 272.6±68.7 N and mixed manufacturer: 299.2±89.0 N | Lever-out test |