Table 7.

Summary of the in-vivo wear studies (All the values listed below are the mean values)

Researcher Test subjects Results Testing technique
Geringer [58] 12 explants of DMC Convex: 36.7±29.1 mm3/year In-vivo test (CMM)
Concave: 22.1±39.2 mm3/year
Total wear rate: 53.9±50.3 mm3/year
Imbert [22] 8 explants of DMC Convex: 29.72 mm3/year In-vivo test (CMM)
Concave: 15.96 mm3/year
Boyer [23] 35 explants of DMC Convex: 25 mm3/year In-vivo test (Light Surface Scanner)
Concave: 12 mm3/year
Adam [24] 40 retrieved DM liners convex: 28.9 mm3/year In-vivo test (Surface analysis)
concave: 25.5 mm3/year
total wear rate: 54.3 mm3 year
Deckard [25] 34 implants of DMC Mean volumetric wear: Martell methodology at regular postoperative intervals
783 mm3/year at 1 year
555 mm3/year at 2 years
104 mm3/year at 5 years
Leonard [26] 129 retrieved DM liners The lever-out force for same manufacturer: 272.6±68.7 N and mixed manufacturer: 299.2±89.0 N Lever-out test