Table 7.
Summary of the in-vivo wear studies (All the values listed below are the mean values)
Researcher | Test subjects | Results | Testing technique |
---|---|---|---|
Geringer [58] | 12 explants of DMC | Convex: 36.7±29.1 mm3/year | In-vivo test (CMM) |
Concave: 22.1±39.2 mm3/year | |||
Total wear rate: 53.9±50.3 mm3/year | |||
Imbert [22] | 8 explants of DMC | Convex: 29.72 mm3/year | In-vivo test (CMM) |
Concave: 15.96 mm3/year | |||
Boyer [23] | 35 explants of DMC | Convex: 25 mm3/year | In-vivo test (Light Surface Scanner) |
Concave: 12 mm3/year | |||
Adam [24] | 40 retrieved DM liners | convex: 28.9 mm3/year | In-vivo test (Surface analysis) |
concave: 25.5 mm3/year | |||
total wear rate: 54.3 mm3 year | |||
Deckard [25] | 34 implants of DMC | Mean volumetric wear: | Martell methodology at regular postoperative intervals |
783 mm3/year at 1 year | |||
555 mm3/year at 2 years | |||
104 mm3/year at 5 years | |||
Leonard [26] | 129 retrieved DM liners | The lever-out force for same manufacturer: 272.6±68.7 N and mixed manufacturer: 299.2±89.0 N | Lever-out test |