Table 1.

The variation of sensitivity and specificity of the in-house antigen ELISA system depending on Ct value and storage condition as compared to RT-qPCR result (n = 339)

  ELISA +/ RT-qPCR + ELISA -/ RT-qPCR- Sensitivity
(95% CI)
Specificity
(95% CI)
PPV
(95% CI)
NPV
(95% CI)
With All Ct Value
Overall 91/147 190/192 61.9
(53.5, 69.8)
99.0
(96.3, 99.9)
97.8
(92.4, 99.7)
77.2
(71.5, 82.3)
4°C followed by −80°C 34/63 53/53 54.0
(40.9, 66.6)
100.0
(93.3, 100)
100
(89.7, 100)
64.6
(53.3, 74.9)
−20°C 4/9 22/22 44.4
(13.7, 48)
100.0
(84.6, 100)
100
(39.8, 100)
81.5
(61.9, 93.7)
−80°C 53/75 115/117 70.7
(59, 80.6)
98.3
(94, 99.8)
96.4
(87.5, 99.6)
83.9
(76.7, 89.7)
Ct ≤35
Overall 80/115 190/192 69.6
(60.3, 77.8)
99.0
(96.3, 99.9)
97.6
(91.5, 99.7)
84.4
(79, 88.9)
4°C followed by −80°C 34/62 53/53 54.8
(41.7, 67.5)
100.0
(93.3, 100)
100
(89.7, 100)
65.4
(54, 75.5)
−20°C 3/5 22/22 60.0
(14.7, 94.7)
100.0
(84.6, 100)
100
(29.2, 100)
91.7
(73.0, 99.0)
−80°C 43/48 115/117 89.6
(77.3, 96.5)
98.3
(94, 99.8)
95.6
(84.9, 99.5)
95.8
(90.5, 98.6)
Ct ≤30
Overall 49/65 190/192 75.4
(63.1, 85.2)
99.0
(96.3, 99.9)
96.1
(86.5, 99.5)
92.2
(87.7, 95.5)
4°C followed by −80°C 22/36 53/53 61.1
(43.5, 76.9)
100.0
(93.3, 100)
100
(84.6, 100)
79.1
(67.4, 88.1)
−20°C 1/1 22/22 100.0
(2.5, 100)
100.0
(84.6, 100)
100.0
(2.5, 100)
100.0
(84.6, 100)
−80°C 26/28 115/117 92.9
(76.5, 99.1)
98.3
(94, 99.8)
92.9
(76.5, 99.1)
98.3
(94, 99.8)

Note: AUC: Area under curve; 95% CI: 95% Confidence interval. Cohen’s Kappa test was used to evaluate the test agreement.