Table 3.
Figure | Scanner Model | Acquisition and Reconstruction | Contrast Agent(s) | Voxel Size (Resolution) | Reported Dose |
---|---|---|---|---|---|
| |||||
1 | CosmoScan FX Rigaku Corporation | - 2 min. at 90 kVp, 88 μA, FOV 45 mm - CosmoScan Database software |
N/A | 90 μm | Not reported |
2 | - Custom-built system within a refurbished clinical CT gantry - Dexela 2923 MAM EID, Perkin Elmer Inc. - L10951 source, Hamamatsu Photonics K. K. |
- 5 min. at 60 kVp, 50 W - 11.7 ms exposure per projection - 10 cardiac, 4 respiratory phases - Intrinsic gating - Motion compensated reconstruction [10] - Post-reconstruction denoising [11] |
ExiTron nano 12,000 (Ba based), nanoPET Pharma GmbH | <75 μm (10% MTF: 7.5 lp/mm) | 5 Gy (0.5–2 Gy results also demonstrated) |
3 | - Custom-built ex vivo scanner - Rotating specimen geometry - XCounter Thor PCD, Direct Conversion AB - L9181-02 source, Hamamatsu Photonics K. K. |
- 2 h at 80 kVp, 0.2 mA - Helical acquisition with 2.5 cm vertical translation (1070 projections, 1070°) - Split Bregman algebraic reconstruction regularized with rank-sparse kernel regression [12] |
BriteVu (Ba based), Scarlet Imaging | 38 μm (10% MTF: 6.5–7.1 lp/mm) | N/A (ex vivo) |
4 | - Custom dual-source, dual-energy in vivo scanner - Vertical subject geometry - Dexela 1512CL EID (CsI), Perkin Elmer Inc. - G-297 sources, Varex Imaging |
- Chain 1: 40 kVp, 50 mA, 25 ms - Chain 2: 80 kVp, 40 mA, 10 ms - Circular scan (360 projections, 360°) - Analytical reconstruction - Post-reconstruction denoising with joint bilateral filtration [13] |
- Iodine liposomes [14] – 15 nm AuroVist gold nanoparticles |
63 μm (10% MTF: 3.4 lp/mm) | 57 mGy |
5A | Same as Fig. 4 | - Chain 1: 40 kVp, 50 mA, 25 ms - Chain 2: 50 kVp, 80 mA, 12.5 ms - Circular scan (720 projections, 360°) - Split Bregman algebraic reconstruction regularized with rank-sparse kernel regression [12] |
- Iodine liposomes [14] - Gadolinium liposomes [15] |
123 μm (10% MTF: 3.4 lp/mm) | 162 mGy |
5B | - Custom-built in vivo scanner - Vertical subject geometry - SANTIS 0804 ME prototype PCD, Dectris AG - G-297 source, Varex Imaging |
- 3 min. at 80 KVp, 2 mA, 200 ms - Helical acquisition, 1.25 cm vertical translation (900 projections, 1070°) - Energy thresholds: 25, 34, 50, 60 keV - Split Bregman algebraic reconstruction regularized with rank-sparse kernel regression [12] |
- Iodine liposomes [14] - Gadolinium liposomes [15] |
123 μm (10% MTF: 3.5 lp/mm) | 43 mGy |
6 | Same as Fig. 5B | - 90 sec. at 80 KVp, 5 mA, 10 ms - Helical acquisition, 1.25 cm vertical translation (9000 projections, 1070°) - 10 cardiac phases, retrospective gating - Energy thresholds: 25, 34, 40, 55 keV - Split Bregman algebraic reconstruction regularized with rank-sparse kernel regression [12] |
- Iodine liposomes [14] – 15 nm AuroVist gold nanoparticles |
123 μm (10% MTF: 2.8–3.0 lp/mm depending on energy threshold) | 190 mGy |
7 | Similar to Fig. 4 | - Iodine liposomes [14] | |||
8 | - Edge-illumination XPC imaging - Pixirad-2 PCD [16] - MicroMax-007 HF, Rigaku source (Mo anode) - 30 μm Mo source filter |
- 18 h at 40 kVp, 30 mA, 1 s - 1441 projection angles over 360° - 5 phase steps, 4 dithering steps per angle - Reconstruction with filtered backprojection (Hilbert filter for phase) - Sample translation during scanning to reduce ring artifacts |
N/A | (19 μm FWHM) | Not measured (lower dose 300 mGy scan also presented) |
9 | Same CT data as in Fig. 6 | - Only 25 keV threshold data used for 4D CNN training - 10 cardiac phases, retrospective gating - Split Bregman algebraic reconstruction regularized with rank-sparse kernel regression [12] (Fig. 9B; CNN training labels; 9000 projections over 1070°) - Unregularized algebraic reconstruction (Fig. 9C; CNN training inputs; subsampled to 2250 projections over 1070°) |
123 μm (10% MTF: 2.7 lp/mm, 4D CNN output, Fig. 9D) | ||
10 | - Non-contrast scans: TomoScope Duo commercial scanner, formerly CT Imaging GmbH - Contrast-enhanced scans: InSyTe μCT commercial scanner, Trifoil Imaging |
- Non-contrast scans: 90 sec. per scan, 65 kVp, 1 mA; 720 projections over 360° - Contrast-enhanced scans: 75 kVp, 230 ms; 207 projections over 360° |
- No contrast or ExiTron nano 6000 (Ba based), nanoPET Pharma GmbH | - Non-contrast scans: 80 μm spatial resolution reported - Contrast scans: 280 μm voxels - Segmentations processed with 240 μm voxels |
Not reported |