Results for air kerma and air kerma rate measurements for the lateral projection, with the workers positioned at distances of 30, 60, 100, and 200 cm

Device Source-worker
distance (cm)
Air kerma
(µGy)
Air kerma rate
(µGy/s)
Philips BV Pulsera 50 (tube side) 13.13±0.65 0.87±0.04
Philips BV Pulsera 50 (intensifier side) 0.55±0.01 0.04±0.00
Philips BV Pulsera 60 0.52±0.01 0.03±0.00
Siemens Siremobil Compact L (B) 50 (tube side) 7.02±0.35 0.47±0.01
Siemens Siremobil Compact L (B) 50 (intensifier side) 0.50±0.01 0.03±0.00
Siemens Siremobil Compact L (B) 60 0.70±0.01 0.05±0.00
Siemens Siremobil Compact L (C) 50 (tube side) 12.57±0.62 0.84±0.04
Siemens Siremobil Compact L (C) 50 (intensifier side) 0.52±0.01 0.03±0.00
Siemens Siremobil Compact L (C) 60 0.68±0.01 0.04±0.00
Siemens Siremobil Compact L (C) 50 (height: 100 cm) (tube side) 9.43±0.48 0.63±0.00
Siemens Siremobil Compact L (C) 50 (height: 100 cm) (intensifier side) 4.51±0.22 0.30±0.00