. 1999 Mar;37(3):611–614. doi: 10.1128/jcm.37.3.611-614.1999

TABLE 1.

Clinical evaluation of CWP1 as a diagnostic marker with the TechLab Giardia Testa

TechLab Giardia Test parameter Result for:
Study 1 (Sacred Heart Medical Center, Spokane, Wash.) Study 2 (Dekalb Medical Center, Decatur, Ga.) Study 3 (VRL, San Antonio, Tex.)
Total no. of stools 114 221 200
Total no. positive 31 46 16
Total no. negative 79 175 180
Total no. of discrepancies 4 0 4
% Sensitivity 91.4 100.0 100.0
% Specificity 100.0 100.0 97.8
Positive predictive value (%) 100.0 100.0 80.0
Negative predictive value (%) 96.3 100.0 100.0
% Correlation 97.4 100.0 98.0
a

The ELISA was performed as instructed by the manufacturer. O&P was done by in-house methods implemented at each study site. All stool specimens were preserved in 10% buffered formalin.