Skip to main content
. 2021 May 14;23:68–77. doi: 10.1016/j.omtm.2021.05.003

Table 3.

Accuracy and precision of the one-step RT-ddPCR method in quantifying the potency of a gene-repression vector (linearity)

Expected % relative potency Measured % relative potency Average measured % relative potency % Recovery
150 162 164.0 109.3
166
125 132 130.5 104.4
129
75 72.9 72.8 97.0
72.6
50 48.9 49.5 99.0
50.1
Accuracy (average % recovery) 98.0
Precision (%CV) 1.4