Skip to main content
. 1999 Mar;37(3):792–795. doi: 10.1128/jcm.37.3.792-795.1999

TABLE 4.

Ultrasensitive quantification of samples not detecteda by Quantiplex HIV RNA 2.0 assay (bDNA)

Sample Ultrasensitive Amplicor HIV-1 Monitor test (HIV-1 RNA copies/ml) Quantiplex HIV RNA 2.0 assay (bDNA)
RL Ub RL U above STD Dc
1 190 0.862 0.084
2 350 0.750 0.241
3 470 0.553 0.149
4 1,200 0.858 0.003
5 1,400 0.950 0.167
6 1,500 0.648 0.178
7 1,600 0.709 0.174
8 1,900 0.919 0.331
9 2,200 0.868 0.238
10 2,300 0.742 0.111
a

The detection limit of the Quantiplex HIV RNA 2.0 assay (bDNA) is 500 copies/ml. 

b

The mean RL of the sample in the assay. 

c

Calculated by using sample RL − STD D RL.