Flap thickness obtained with manual and femtosecond methods for corneal flap creation in clinical LASIK studies.

Study Number of Eyes Method of Flap Creation Intended Thickness [μm] Method for Flap Thickness Measurement (Device Name) Measured Central Flap Thickness ±SD [μm] Timing of Flap Thickness Measurement
Manual microkeratomes
Kezirian and Stonecipher 2004 [116] 126 Carriazo-Barraquer (Moria) 130 USP (DGH Pachette 50/60 KHz) 153 ± 26 After lifting the flap (total thickness − thickness after flap lift)
Kezirian and Stonecipher 2004 [116] 143 Hansatome (Bausch & Lomb) 180 USP (DGH Pachette 50/60 KHz) 156 ± 29 After lifting the flap (total thickness − thickness after flap lift)
Choudhri et al., 2005 [32] 138 Hansatome (Bausch & Lomb) 160 USP (American Surgical Instruments Corneal Gauge) 124.1 ± 17.4 After lifting the flap (total thickness − thickness after flap lift)
Choudhri et al., 2005 [32] 112 Hansatome (Bausch & Lomb) 180 USP (American Surgical Instruments Corneal Gauge) 142.3 ± 19.6 After lifting the flap (total thickness − thickness after flap lift)
Duffey 2005 [23] 42 LSK-1 (Moria) 100 USP (DGH Pachette) 107.0 ± 14.0 After lifting the flap (total thickness − thickness after flap lift)
Talamo et al., 2006 [35] 100 LSK-1 (Moria) 160 USP (DGH Pachette II) 130.0 ± 19.0 After lifting the flap (total thickness − thickness after flap lift)
Talamo et al., 2006 [35] 135 M2 (Moria) 130 USP (DGH Pachette II) 142.0 ± 24 After lifting the flap (total thickness − thickness after flap lift)
Javaloy et al., 2007 [40] 100 M2 (Moria) 160 confocal microscopy (ASL 165a) 149.1 ± 14.0 1 and 3 months postop
Patel el al., 2007 [26] 21 Hansatome (Bausch & Lomb) 180 confocal microscopy (Nidek ConfoScan 3 or 4) 138.0 ± 22 1 month postop
Huhtala et al., 2007 [34] 300 M2 (Moria) 120 USP (Cilco) 115.4 ± 12.5 After lifting the flap (total thickness − thickness after flap lift)
Alió and Piñero 2008 [6] 22 M2 (Moria) 110 high-frequency USP (Ultralink Artemis) 117.5 ± 7.8 1 month postop
Alió and Piñero 2008 [6] 22 Carriazo-Pendular Microkeratome (Schwind) 110 high-frequency USP (Ultralink Artemis) 118.1 ± 8.3 1 month postop
Hamilton et al., 2008 [30] 32 One Use (Moria) 130 USP (N/A) 117.0 ± 16.0 After lifting the flap (total thickness − thickness after flap lift)
Rosa et al., 2009 [20] 20 Hansatome Zero Compression (Bausch & Lomb) 160 USP (Sonogage Corneo-Gage Plus) 149.1 ± 24.9 After lifting the flap (total thickness − thickness after flap lift); in 1of 4 group 20 min postop
Rosa et al., 2009 [20] 20 Zyoptix XP (Bausch & Lomb) 120 USP (Sonogage Corneo-Gage Plus) 124.7 ± 23.8 After lifting the flap (total thickness − thickness after flap lift); in 1of 4 group 20 min postop
Salomão et al., 2009 [31] 70 Hansatome (Bausch & Lomb) 180 USP (N/A) 131.0 ± 25 After lifting the flap (total thickness − thickness after flap lift)
Ahn et al., 2011 [27] 52 M2 (Moria) 130 OCT (Optovue RTVue FD-OCT) 126.0 ± 19.9 ** 2 months postop
Yao et al., 2011 [37] 38 M3 (Moria) 110 OCT (Carl Zeiss Visante) 112.2 ± 5.4 1 week–6 months postop
Zhang et al., 2014 [117] 50 M2 (Moria) 110 USP (DGH 550) 133.0 ± 13.9 After lifting the flap (total thickness − thickness after flap lift)
Zhang et al., 2014 [33] 60 One Use-Plus SBK (Moria) 110 USP (DGH 550) 109.4 ± 6.8 After lifting the flap (total thickness − thickness after flap lift)
Abdelwahab and Elfayoumi 2016 [52] 500 One Use-Plus SBK (Moria) 100 USP (DGH 55 Pachmate) 102.0 ± 6.1 After lifting the flap (total thickness − thickness after flap lift)
Torky et al., 2017 [118] 23 M2 (Moria) N/A USP (Tomey SP 100) 104.6 ± 20.1 After lifting the flap (total thickness − thickness after flap lift)
Karabela et al., 2017 [119] 72 M2 (Moria) 120 USP (Nidek Echoscan US-1800) 134.2 ± 19.9 After lifting the flap (total thickness − thickness after flap lift)
Eldaly et al., 2019 [28] 22 M2 (Moria) 100–110 OCT (Heidelberg Engineering Spectralis) 125.8 ± 5.8 1 month postop
Eldaly et al., 2019 [28] 22 M2 (Moria) 130 OCT (Heidelberg Engineering Spectralis) 146.9 ± 13.6 1 month postop
Femtosecond laser
Kezirian and Stonecipher 2004 [116] 106 IntraLase S3 (Abbott Medical Optics) 130 USP (DGH Pachette 50/60 KHz) 114 ± 14 After lifting the flap (total thickness − thickness after flap lift)
Binder 2004 [120] 34 IntraLase S3 (Abbott Medical Optics) 110 USP (Sonnogage Cornea Scan II 5) 125.0 ± 12.0 After lifting the flap (total thickness − thickness after flap lift)
Binder 2004 [120] 22 IntraLase S3 (Abbott Medical Optics) 120 USP (Sonnogage Cornea Scan II 5) 122.4 ± 11.9 After lifting the flap (total thickness − thickness after flap lift)
Binder 2004 [120] 21 IntraLase S3 (Abbott Medical Optics) 130 USP (Sonnogage Cornea Scan II 5) 128.7 ± 16.6 After lifting the flap (total thickness − thickness after flap lift)
Binder 2004 [120] 26 IntraLase S3 (Abbott Medical Optics) 140 USP (Sonnogage Cornea Scan II 5) 132.5 ± 18.5 After lifting the flap (total thickness − thickness after flap lift)
Talamo et al., 2006 [35] 99 IntraLase FS (Abbott Medical Optics) 110 USP (DGH Pachette II) 119.0 ± 12 After lifting the flap (total thickness − thickness after flap lift)
Javaloy et al., 2007 [40] 100 IntraLase FS (Abbott Medical Optics) 120 confocal microscopy (ASL 165a) 130.1 ± 1.7 1 and 3 months postop
Patel el al., 2007 [26] 21 IntraLase 15- kHz (Abbott Medical Optics) 120 confocal microscopy (Nidek ConfoScan 3 or 4) 143.0 ± 16 1 month postop
Alió and Piñero 2008 [6] 22 IntraLase 30- kHz (Abbott Medical Optics) 110 high-frequency USP (Ultralink Artemis) 116.0 ± 6.2 1 month postop
Hamilton et al., 2008 [30] 32 IntraLase 60- kHz (Abbott Medical Optics) 110–120 USP (N/A) 120.0 ± 13 After lifting the flap (total thickness − thickness after flap lift)
Sutton and Hodge 2008 [102] 838 IntraLase 15- kHz (Abbott Medical Optics) 105 USP (Sonogage Corneo-Gage Plus) 116.8 ± 10.8 After lifting the flap (total thickness − thickness after flap lift)
Sutton and Hodge 2008 [102] 162 IntraLase 30- kHz (Abbott Medical Optics) 115 USP (Sonogage Corneo-Gage Plus) 114.0 ± 9.8 After lifting the flap (total thickness − thickness after flap lift)
Rosa et al., 2009 [20] 20 IntraLase FS 60 kHz (Abbott Medical Optics) 120 USP (Sonogage Corneo-Gage Plus) 115.5 ± 12.5 After lifting the flap (total thickness − thickness after flap lift); in 1of 4 group 20 min postop
Salomão et al., 2009 [31] 113 IntraLase 30- or 60- kHz (Abbott Medical Optics) 100–110 USP (N/A) 131.0 ± 25 After lifting the flap (total thickness − thickness after flap lift)
Ahn et al., 2011 [27] 50 IntraLase 110 OCT (Optovue RTVue FD-OCT) 130.3 ± 13.2 ** 2 months postop
Ahn et al., 2011 [27] 40 VisuMax (Carl Zeiss Meditec) 110 OCT (Optovue RTVue FD-OCT) 133.9 ± 13.9 ** 2 months postop
Ahn et al., 2011 [27] 64 Femto LDV (Ziemer) 110 OCT (Optovue RTVue FD-OCT) 105.8 ± 8.2 ** 2 months postop
Yao et al., 2011 [37] 25 VisuMax (Carl Zeiss Meditec) 100 OCT (Carl Zeiss Visante) 114.2 ± 6.9 1 week–6 months postop
Kim et al., 2011 [36] 19 IntraLase (Abbott Medical Optics) 110 OCT (Carl Zeiss Visante) 115.2 ± 5.0 1 week postop
Kim et al., 2011 [36] 7 IntraLase (Abbott Medical Optics) 120 OCT (Carl Zeiss Visante) 121.9 ± 5.8 1 week postop
Kim et al., 2011 [36] 9 IntraLase (Abbott Medical Optics) 130 OCT (Carl Zeiss Visante) 134.4 ± 5.0 1 week postop
Zhang et al., 2014 [117] 72 FS200 (Alcon) 110 USP (DGH 550) 105.5 ± 5.9 After lifting the flap (total thickness − thickness after flap lift)
Zhang et al., 2014 [33] 60 FS200 (Alcon) 110 USP (DGH 550) 112.7 ± 7.2 After lifting the flap (total thickness − thickness after flap lift)
Zheng et al., 2015 [24] 200 FS200 (Alcon) 110 OCT (OptoVue RTVue FD-OCT) 105.4 ± 3.4 1 week postop
Zheng et al., 2015 [24] 200 VisuMax (Carl Zeiss Meditec) 110 OCT (OptoVue RTVue FD-OCT) 110.8 ± 3.9 1 week postop
Liu et al., 2016 [38] 200 FS200 (Alcon) 110 OCT (OptoVue RTVue FD-OCT) 105.4 ± 4.5 1 week postop
Liu et al., 2016 [38] 200 IntraLase FS60 (Abbott Medical Optics) 110 OCT (OptoVue RTVue FD-OCT) 109.2 ± 11.6 1 week postop
Torky et al., 2017 [118] 26 Visumax FSL (Carl Zeiss Meditec) 100 USP (Tomey SP 100) 100.1 ± 16.1 After lifting the flap (total thickness − thickness after flap lift)
Eldaly et al., 2019 [28] 25 FS200 (Alcon) 100–110 OCT (Heidelberg Engineering Spectralis) 114.6 ± 6.1 1 month postop
Eldaly et al., 2019 [28] 24 FS200 (Alcon) 130 OCT (Heidelberg Engineering Spectralis) 140.8 ± 13.8 1 month postop
Parafita-Fernandez et al., 2020 [39] 44 FS200 (Alcon) 120 OCT (Zeiss Cirrus HD-OCT 5000) 132.2 ± 8.1 1 day–3 months postop

Abbreviations: OCT—optical coherence tomography, SD—standard deviation, USP—ultrasound pachymetry; ** central thickness in horizontal plane.